EchoNavigator R5.0
K253614Philips Medical Systems Nederland B.V. · cleared 2026-03-17 · product code QIH · Radiology
Premarket evidence — what FDA accepted
source quote (p.6)
“EchoNavigator is a Software Medical Device that assists the interventionalist and surgeon with image guidance during treatment of cardiovascular disease for which the procedure uses both live X-ray and live Echo guidance.”
source quote (p.7)
“DeviceGuide offers real-time automatic device detection, tracking and 3D pose-estimation in live Echo and live X-ray enabled by Artificial Intelligence which is built with Deep Learning technology.”
source quote (p.11)
“To ensure independence between development and testing, testing was performed after algorithm development was completed and the algorithm was frozen.”
source quote (p.10)
“Privacy and Security requirements and mitigations have been implemented.”
Validation studies (2)
Standalone
n=72 patients
endpoints: positional accuracy; trajectory/orientation accuracy; algorithm latency
standards: IEC 62304, IEC 62366-1, IEC 82304-1, ISO 14971, ISO 15223-1, ISO 20417:2021, UL ANSI 2900-1, UL ANSI 2900-2-1, IEC 81001-5-1
Reader study (MRMC)
sample size not stated
endpoints: physician assessment of algorithm output sufficiency
Reported performance (3 observations)
source quote (p.11)
“Overall performance met predefined specification limits, including ≤5 mm positional accuracy and ≤10° trajectory/orientation accuracy.”
source quote (p.11)
“Overall performance met predefined specification limits, including ≤5 mm positional accuracy and ≤10° trajectory/orientation accuracy.”
source quote (p.11)
“Algorithm latency met predefined specification, including detection and localization of the therapy device within 100ms of reception of the echo image.”
Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.
Predicate network
Postmarket — what happened after clearance
Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device). Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.
Reimbursement — how devices like this got paid
Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).