BriefCase-Triage: CARE Multi-Triage CT for Pneumothorax; Pericardial effusion; Large aortic aneurysm; Shoulder fracture or dislocation device
K253578Aidoc Medical , Ltd. · cleared 2026-02-26 · product code QAS · Radiology
Premarket evidence — what FDA accepted
source quote (p.5)
“BriefCase-Triage: CARE Multi-Triage CT for Pneumothorax; Pericardial effusion; Large aortic aneurysm; Shoulder fracture or dislocation device is a radiological computer-assisted triage and notification software device.”
source quote (p.10)
“Multi-triage module, locked artificial intelligence algorithm fine tuned from a foundation model.”
source quote (p.10)
“Multi-triage module, locked artificial intelligence algorithm fine tuned from a foundation model.”
source quote (p.16)
“Cybersecurity has been incorporated into the software development lifecycle in alignment with Section 524B of the FD&C Act and FDA cybersecurity guidance.”
Validation studies (1)
Retrospective clinical
n=772 scans · 6 site(s)
endpoints: area under the curve (AUC) > 0.95 lower bound 95% confidence interval for the finding level receiver operating characteristic (ROC) curve; sensitivity; specificity; BriefCase time-to-notification compared to the predicate device
Reported performance (16 observations)
source quote (p.11)
“94.8% 89.5%-97.9%”
source quote (p.11)
“95.9% 92.0%-98.8%”
source quote (p.11)
“98.9 97.8-99.7”
source quote (p.11)
“98.9 97.8-99.7”
source quote (p.11)
“94.8% 89.5%-97.9%”
source quote (p.11)
“95.9% 92.0%-98.8%”
source quote (p.11)
“99.1 98.0-99.8”
source quote (p.11)
“96.4% 91.7%-98.8%”
source quote (p.11)
“96.5% 92.9%-99.2%”
source quote (p.11)
“99.5 98.9-99.9”
source quote (p.11)
“97.1% 92.7%-99.2%”
source quote (p.11)
“97.2% 96.2%-100.0%”
source quote (p.11)
“99.9 99.7-100”
source quote (p.11)
“97.8% 93.7%-99.5%”
source quote (p.11)
“99.3% 91.3%-98.5%”
source quote (p.11)
“The BriefCase-Triage: CARE Multi-Triage CT for Pneumothorax; Pericardial effusion; Large aortic aneurysm; Shoulder fracture or dislocation mean time-to-notification for the subject device triage was 49.9 seconds (95% CI: 46.4-53.5).”
Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.
Predicate network
Postmarket — what happened after clearance
Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device). Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.
Reimbursement — how devices like this got paid
Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).