AI-Rad Companion Brain MR
K253057Siemens Healthcare GmbH · cleared 2026-01-22 · product code QIH · Radiology
Premarket evidence — what FDA accepted
source quote (p.4)
“Al-Rad Companion Brain MR is a post-processing image analysis software that assists clinicians in viewing, analyzing, and evaluating MR brain images.”
source quote (p.7)
“AI-Rad Companion Brain MR runs two distinct and independent algorithms for Brain Morphometry analysis and White Matter Hyperintensities (WMH) segmentation, respectively.”
source quote (p.14)
“Siemens Healthineers adheres to the cybersecurity recommendations as defined the FDA Guidance “Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions” (September 2023) by implementing a process of preventing unauthorized access, modifications, misuse or denial of use, or the unauthorized use of information that is stored, accessed, or transferred from a medical device to an external recipient.”
Validation studies (2)
Standalone
n=100 patients
endpoints: Pearson correlation coefficient; interclass correlation coefficient; Dice score; F1-score
standards: IEC 62366-1, ISO 14971, IEC 62304, PS 3.1 – 3.20, 15223-1, 82304-1
Standalone
n=165 patients
endpoints: Pearson correlation coefficient; Dice score; F1-score
standards: IEC 62366-1, ISO 14971, IEC 62304, PS 3.1 – 3.20, 15223-1, 82304-1
Reported performance (4 observations)
source quote (p.14)
“The segmentation accuracy of the WMH reaches a Dice score of 0.60.”
source quote (p.14)
“The F1-score of WMH detection is 0.67.”
source quote (p.15)
“The segmentation accuracy of the new or enlarged WMH reaches an average Dice score of 0.59.”
source quote (p.15)
“The average F1-score of new or enlarged WMH detection is 0.71.”
Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.