BriefCase-Triage: CARE Multi-triage CT Body
K252970Aidoc Medical , Ltd. · cleared 2026-01-07 · product code QAS · Radiology
Premarket evidence — what FDA accepted
source quote (p.5)
“Briefcase-Triage is a radiological computer-assisted triage and notification software device.”
source quote (p.11)
“Multi-triage module, locked artificial intelligence algorithm fine tuned from a foundation model.”
source quote (p.11)
“Multi-triage module, locked artificial intelligence algorithm fine tuned from a foundation model.”
source quote (p.27)
“Cybersecurity has been incorporated into the software development lifecycle in alignment with Section 524B of the FD&C Act and FDA cybersecurity guidance.”
Validation studies (1)
Standalone
n=1,769 scans · 6 site(s)
endpoints: pre-specified standalone performance goal (PG) of area under the curve (AUC) > 0.95 for the finding level receiver operating characteristic (ROC) curve; sensitivity; specificity; BriefCase time-to-notification compared to the predicate device; Positive Predictive Value (PPV); Negative Predictive Value (NPV); Positive Likelihood Ratio (PLR); Negative Likelihood Ratio (NLR)
Reported performance (4 observations)
source quote (p.12)
“98.6% 95.1%-99.8%”
source quote (p.12)
“98.5% 94.7%-99.8%”
source quote (p.12)
“99.9 99.7-100”
source quote (p.14)
“The Briefcase-Triage mean time-to-notification for the subject device triage was 45 seconds (95% CI: 43.4-46.5).”
Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.
Predicate network
Postmarket — what happened after clearance
- re_clearance
The FDA AI/ML device list shows a newer 510(k) K251195 (decision 2026-01-27) from Aidoc Medical , Ltd. for a matching device line ("BriefCase-Triage") — a new clearance for the same line is a change event.
first seen 2026-07-08 · k_number:K251195
Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device). Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.