Monaco RTP System (6.3)
K252002Elekta Solutions AB · cleared 2026-02-19 · product code MUJ · Radiology
Premarket evidence — what FDA accepted
source quote (p.5)
“Monaco is programmed using C, C++ and C# computer programming languages. Monaco runs on Windows operating system and off-the-shelf computer server/hardware.”
source quote (p.7)
“The Segmentation Component utilizes machine-learning based models to automatically segment MR image sets introduced in subject device.”
source quote (p.8)
“IEC 81001-5-1”
Validation studies (1)
Standalone
sample size not stated
endpoints: Average Hausdorff Distance (AVD); DICE Similarity Coefficient (DICE); AUC; Absolute Volume Difference (AVD)
standards: ISO 13485, ISO 14971, IEC 62304, IEC 62083, IEC 82304-1, IEC 61217, IEC 62366-1, IEC 81001-5-1, ISO 15223-1, ISO 20417
Reported performance (0 observations)
FDA source did not state a quantitative performance metric — non-reporting is itself the signal.
Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.
Predicate network
Postmarket — what happened after clearance
- recall_reason_pattern
Software/algorithm-related recall in product code MUJ (Philips Medical Systems (Cleveland) Inc, initiated 2025-08-05): "Due to a software issue, there is a potential image error of the Region of Interest for expansion/contraction for HFP (Head First Prone), FFS (Feet First Supine) and FFP (Feet Firs" Recalling firm is another firm in the same product code.
first seen 2026-07-08 · recall res_event_number:97049
- recall_reason_pattern
Software/algorithm-related recall in product code MUJ (Philips Medical Systems (Cleveland) Inc, initiated 2025-07-17): "Due to software issue, Radiation Therapy Planning system may provide incorrect dataset calculations when performing the "Stopping Power Ratio" (SPR) ," Recalling firm is another firm in the same product code.
first seen 2026-07-08 · recall res_event_number:97309
Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device). Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.
Reimbursement — how devices like this got paid