Cartesion Prime (PCD-1000A/3) V10.21

K251370

Canon Medical Systems Corporation · cleared 2025-12-01 · product code KPS · Radiology

Premarket evidence — what FDA accepted

Device typehardware with ml
source quote (p.6)
The Cartesion Prime (PCD-1000A/3) V10.21 combines a high-end CT and a high-throughput PET designed to acquire CT, PET and fusion images.
Algorithmdeep learning artificial neural network
source quote (p.6)
The subject device incorporates the latest reconstruction technology, AiCE-i for PET (Advanced Intelligent Clear-IQ Engine- integrated), intended to improve image quality and reduce image noise for FDG whole body data by employing deep learning artificial neural network methods which can more fully explore the statistical properties of the signal and noise of PET data.
Adaptive (vs locked)FDA source did not state this
PCCPNo
Cybersecurity addressedYes
source quote (p.10)
Cybersecurity documentation, per the FDA guidance document “Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions”, issued on September 27, 2023, was included in this submission.

Validation studies (3)

Reader study (MRMC)

n=10 cases

endpoints: overall image quality; image sharpness; image noise

standards: IEC60601-1, IEC60601-1-2, IEC60601-1-3, IEC60601-1-6, IEC60601-2-28, IEC60601-2-44, IEC60825-1, IEC62304, IEC81001-5-1, IEC62366-1, NEMA XR-25, NEMA XR-26, NEMA XR-29, NEMA NU-2

Standalone

n=11 patients

endpoints: quantified outputs on high uptake regions (e.g., lesions)

standards: IEC60601-1, IEC60601-1-2, IEC60601-1-3, IEC60601-1-6, IEC60601-2-28, IEC60601-2-44, IEC60825-1, IEC62304, IEC81001-5-1, IEC62366-1, NEMA XR-25, NEMA XR-26, NEMA XR-29, NEMA NU-2

Reader study (MRMC)

n=10 patients

endpoints: diagnostic quality; motion correction

standards: IEC60601-1, IEC60601-1-2, IEC60601-1-3, IEC60601-1-6, IEC60601-2-28, IEC60601-2-44, IEC60825-1, IEC62304, IEC81001-5-1, IEC62366-1, NEMA XR-25, NEMA XR-26, NEMA XR-29, NEMA NU-2

Reported performance (0 observations)

FDA source did not state a quantitative performance metric — non-reporting is itself the signal.

Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.

Predicate network

Postmarket — what happened after clearance

34
recalls in product code, 24mo
8
MAUDE reports in code, 12mo
-72%
vs code's own 3-yr baseline
4
drift signals on this device
  • recall_reason_pattern

    Software/algorithm-related recall in product code KPS (GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING 9, Andrey Sakharov Haifa Israel, initiated 2025-12-24): "There is a potential intermittent issue on certain Omni Legend systems that can result in a streaking artifact in the PET clinical scan images. This streaking artifact is most eas" Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:98269

  • recall_reason_pattern

    Software/algorithm-related recall in product code KPS (GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING 4, Hayozma St Tirat Carmel Israel, initiated 2025-06-20): "Unintended radial detector motion may occur during patient setup or during patient scan if system does not have correct version of gantry software installed. Unintended detector mo" Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:97193

  • recall_reason_pattern

    Software/algorithm-related recall in product code KPS (Hermes Medical Solutions AB Strandbergsgatan 16 Stockholm Sweden, initiated 2024-10-31): "Due a potential software/configuration issue that may result is incorrect alignment during reconstructing a SPECT/CT study." Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:95673

  • recall_reason_pattern

    Software/algorithm-related recall in product code KPS (Canon Medical System, USA, INC., initiated 2024-09-17): "When PET-CT system is executing reconstruction, if PET acquisition for another patient is performed (or PET reconstruction for another patient is performed from raw data processing" Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:95471

Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device). Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.

Reimbursement — how devices like this got paid

Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).

RIGOR™ Precedent · public FDA/CMS data · descriptive decision-support, not regulatory or reimbursement advice. Share this page: radar.healthai.com/precedent/device/K251370