Swoop® Portable MR Imaging® System
K251276Hyperfine, Inc. · cleared 2025-05-21 · product code LNH · Radiology
Premarket evidence — what FDA accepted
source quote (p.4)
“The Swoop Portable MR Imaging System is a portable, ultra-low field magnetic resonance imaging device for producing images that display the internal structure of the head where full diagnostic examination is not clinically practical. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis. The Swoop System image reconstruction algorithm utilizes deep learning to provide improved image quality for T1W, T2W, FLAIR, and DWI sequences.”
source quote (p.6)
“The Swoop System image reconstruction algorithm utilizes deep learning to provide improved image quality for T1W, T2W, FLAIR, and DWI sequences.”
source quote (p.8)
“Testing to verify cybersecurity controls and management.”
Validation studies (3)
Retrospective clinical
n=118 patients
endpoints: Normalized mean squared error (NMSE); structural similarity index (SSIM)
Retrospective clinical
n=43 patients · 8 site(s)
endpoints: Contrast-to-Noise Ratio (CNR)
Reader study (MRMC)
n=46 patients · 11 site(s)
endpoints: image quality (noise, sharpness, contrast, geometric fidelity, artifact, overall image quality); consistency of diagnosis
Reported performance (0 observations)
FDA source did not state a quantitative performance metric — non-reporting is itself the signal.
Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.
Predicate network
Postmarket — what happened after clearance
- re_clearance
The FDA AI/ML device list shows a newer 510(k) K253489 (decision 2025-12-12) from Hyperfine, Inc. for a matching device line ("Swoop® Portable MR Imaging® System") — a new clearance for the same line is a change event.
first seen 2026-07-08 · k_number:K253489
- re_clearance
The FDA AI/ML device list shows a newer 510(k) K250236 (decision 2025-05-30) from Hyperfine, Inc. for a matching device line ("Swoop® Portable MR Imaging® System (V2)") — a new clearance for the same line is a change event.
first seen 2026-07-08 · k_number:K250236
- recall_reason_pattern
Software/algorithm-related recall in product code LNH (Philips North America, initiated 2026-04-14): "The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction " Recalling firm is another firm in the same product code.
first seen 2026-07-08 · recall res_event_number:98779
- recall_reason_pattern
Software/algorithm-related recall in product code LNH (Philips North America, initiated 2025-12-03): "The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS)." Recalling firm is another firm in the same product code.
first seen 2026-07-08 · recall res_event_number:98111
Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device). Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.
Reimbursement — how devices like this got paid
Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).