Revolution Vibe

K250941

GE Medical Systems, LLC · cleared 2025-08-01 · product code JAK · Radiology

Premarket evidence — what FDA accepted

Device typehardware with ml
source quote (p.7)
This proposed device Revolution Vibe is a general purpose, premium multi-slice CT Scanning system consisting of a gantry, table, system cabinet, scanner desktop, power distribution unit, and associated accessories. It also includes the Auto ROI enabled by Al which is integrated within the existing SmartPrep workflow for predicting Baseline and monitoring ROI automatically.
AlgorithmDeep Learning Image Reconstruction for noise reduction and Auto ROI enabled by AI for predicting Baseline and monitoring ROI automatically.
source quote (p.7)
Revolution Vibe inherits virtually all of the key technologies from the predicate such as: high tube current (mA) output, 80 cm bore size with Whisper Drive, Deep Learning Image Reconstruction for noise reduction (DLIR K183202/K213999, GSI DLIR K201745), ASIR-V iterative recon, enhanced Extended Field of View (EFOV) reconstruction MaxFOV 2 (K203617), fast rotation speed as fast as 0.23 second/rot (K213715), and spectral imaging capability enabled by ultrafast kilovoltage(kv) switching (K163213), as well as ECG-less cardiac (K233750). It also includes the Auto ROI enabled by Al which is integrated within the existing SmartPrep workflow for predicting Baseline and monitoring ROI automatically.
Adaptive (vs locked)FDA source did not state this
PCCPFDA source did not state this
Cybersecurity addressedFDA source did not state this

Validation studies (3)

Bench

sample size not stated

standards: AAMI/ANSI ES 60601-1, IEC60601-1 Ed. 3.2, 21 CFR Subchapter J, NEMA standards XR 25, NEMA standards XR 26, NEMA standards XR 28, IEC 61223-3-5 ed.2

Retrospective clinical

n=1,341 images

endpoints: success rates exceeding the pre-established acceptance criteria

Reader study (MRMC)

n=30 cases

endpoints: evaluation of image quality; clinical utility

Reported performance (0 observations)

FDA source did not state a quantitative performance metric — non-reporting is itself the signal.

Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.

Predicate network

Postmarket — what happened after clearance

75
recalls in product code, 24mo
192
MAUDE reports in code, 12mo
-22%
vs code's own 3-yr baseline
9
drift signals on this device
  • recall_reason_pattern

    Software/algorithm-related recall in product code JAK (GE Medical Systems, LLC, initiated 2026-03-26): "GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with " Recalling firm matches this device's applicant.

    first seen 2026-07-08 · recall res_event_number:98738

  • recall_reason_pattern

    Software/algorithm-related recall in product code JAK (PHILIPS MEDICAL SYSTEMS, initiated 2026-03-07): "Philips has identified three software issues: 1. During a continuous CT (CCT) scan, there is the potential that the Gantry could remain at the current scan position after pressing" Recalling firm matches this device's applicant.

    first seen 2026-07-08 · recall res_event_number:98588

  • recall_reason_pattern

    Software/algorithm-related recall in product code JAK (PHILIPS MEDICAL SYSTEMS, initiated 2025-09-25): "Issue 1: The potential for unintentional continued gantry/couch movement when a specific button series is used requiring use of manual stop. Issue 2. When performing a helical/Axia" Recalling firm matches this device's applicant.

    first seen 2026-07-08 · recall res_event_number:97699

  • recall_reason_pattern

    Software/algorithm-related recall in product code JAK (Philips Medical Systems Nederland B.V. Veenpluis, initiated 2025-02-28): "Multiple problems identified with the software version leading to various scanning and image issues, and unintended device movement." Recalling firm matches this device's applicant.

    first seen 2026-07-08 · recall res_event_number:96423

  • recall_reason_pattern

    Software/algorithm-related recall in product code JAK (Siemens Medical Solutions USA, Inc, initiated 2025-12-19): "To remove the software applications from certain CT systems as the applications have not received FDA 510(k) clearance." Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:98206

  • recall_reason_pattern

    Software/algorithm-related recall in product code JAK (Philips North America Llc, initiated 2025-05-29): "Devices with affected software may experience two unintended motion issues that may lead to contact between the Gantry or table with the operator or patient, along with additional " Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:97010

  • …and 3 more.

Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device). Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.

Reimbursement — how devices like this got paid

Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).