Methinks NCCT Stroke
K250685Methinks Software S.L. · cleared 2025-06-16 · product code QAS · Radiology
Premarket evidence — what FDA accepted
source quote (p.5)
“Methinks NCCT Stroke is a radiological computer-assisted triage and notification software device. The Methinks NCCT Stroke device is an AI/ML Software as a Medical Device.”
source quote (p.4)
“Methinks NCCT Stroke uses an artificial intelligence algorithm to analyze images and highlight cases with suspected (1) ICH and (2) LVO in the cloud in parallel to the ongoing standard of care image interpretation.”
source quote (p.8)
“Cybersecurity has been considered and addressed as part of the software verification process, in line with Section 524B of the FD&C Act as a Cyber device. Methinks implemented a risk-based cybersecurity strategy including secure design principles, vulnerability assessment, Software Bill of Materials (SBOM), and penetration testing. These activities ensure the software's safety, effectiveness, and robustness against potential cybersecurity threats.”
Validation studies (2)
Retrospective clinical
n=358 cases
endpoints: evaluate the software's performance in identifying NCCT head images containing intracranial hemorrhage (ICH); evaluate the software's performance in identifying Large Vessel Occlusion (LVO) findings
standards: EN ISO 14971:2019, ISO 62304:2015
Reader study (MRMC)
n=335 cases
endpoints: expert non-inferiority for sensitivity; non-expert superiority for sensitivity
Reported performance (7 observations)
source quote (p.9)
“Specifically, ICH performance was observed at Se: 94.7% (95% CI: 89.3% - 97.8%)”
source quote (p.9)
“Sp: 99.5% (95% CI: 97.5% - 99.9%)”
source quote (p.9)
“and LVO was observed at Se: 76.4% (95% CI: 67.3% - 83.9%)”
source quote (p.9)
“and Sp: 91.1% (95% CI: 86.6% - 94.5%).”
source quote (p.9)
“LVO sensitivity (Se) for expert neuroradiologists was 50.0% (40.1% - 59.9%)”
source quote (p.9)
“Time to notification of NCCT-ICH (minutes) 1.43 1.36 1.50”
source quote (p.9)
“Time to notification of NCCT-LVO (minutes) 1.42 1.36 1.48”
Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.
Predicate network
Postmarket — what happened after clearance
Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device). Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.
Reimbursement — how devices like this got paid
Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).