LUMINOS Q.namix T; LUMINOS Q.namix R
K250660Siemens Medical Solutions · cleared 2025-07-14 · product code OWB · Radiology
Premarket evidence — what FDA accepted
source quote (p.4)
“LUMINOS Q.namix T and LUMINOS Q.namix R are devices intended to visualize anatomical structures by converting an X-ray pattern into a visible image. It is a multifunctional, general R/F system, suitable for routine radiography and fluoroscopy examinations, including gastrointestinal- and urogenital examinations and specialist areas like arthrography, angiography and pediatrics.”
source quote (p.9)
“AI-based Auto Cropping”
source quote (p.17)
“Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submission - Guidance for Industry and Food and Drug Administration Staff Document Issued on September 27, 2023”
Validation studies (1)
Retrospective clinical
sample size not stated · 5 site(s)
endpoints: images are acceptable for diagnostic in radiography & fluoroscopy; images and fluorography studies with this detector were qualified for proper diagnosis
standards: ANSI AAMI 60601-1, 2020 Ed. 3.2, IEC 60601-1-2 2020 Ed 4.1, IEC 60601-1-3: Edition 2.2, 2021, IEC 60601-2-28, 2017, IEC 60601-2-54 2018, Edition 1.2, IEC 60601-1-6 2020 Ed 3.2, IEC 62366-1 2020 Ed 1.1, ISO 14971: 2019, IEC 62304 2015, Ed.1.1, IEC 61910-1: 2014, Ed 1.0, NEMA PS 3.1 - 3.20 2023e, ISO EN ISO 10993-1 Fifth edition 2018
Reported performance (0 observations)
FDA source did not state a quantitative performance metric — non-reporting is itself the signal.
Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.
Predicate network
Postmarket — what happened after clearance
- recall_reason_pattern
Software/algorithm-related recall in product code OWB (Siemens Medical Solutions USA, Inc, initiated 2026-05-20): "Potential for loss of movement capability after adjusting the detector lift height to its maximum or minimum limits." Recalling firm matches this device's applicant.
first seen 2026-07-08 · recall res_event_number:99079
- recall_reason_pattern
Software/algorithm-related recall in product code OWB (Siemens Medical Solutions USA, Inc, initiated 2025-12-29): "Ortho images acquired with preset automatic horizontal flip are not flipped and when acquiring single images, there is the possibility that an image of a previous patient may be pr" Recalling firm matches this device's applicant.
first seen 2026-07-08 · recall res_event_number:98244
- recall_reason_pattern
Software/algorithm-related recall in product code OWB (Siemens Medical Solutions USA, Inc, initiated 2025-12-05): "A potential issue associated with LUMINOS Lotus Max and Luminos dRF Max systems with software version VF11 with bucky wall stand for mobile detectors and without ceiling stand .Whe" Recalling firm matches this device's applicant.
first seen 2026-07-08 · recall res_event_number:98230
- recall_reason_pattern
Software/algorithm-related recall in product code OWB (Medtronic Navigation, Inc.-Boxborough, initiated 2026-05-19): "Potential for image artifacts caused by an anomaly in the O-arm O2 Imaging System s detector panel firmware." Recalling firm is another firm in the same product code.
first seen 2026-07-08 · recall res_event_number:99041
- recall_reason_pattern
Software/algorithm-related recall in product code OWB (PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Veenpluis 4-6 Best Netherlands, initiated 2025-12-12): "Six (6) identified software issues which may result in the following: system keeps restarting, AMC triple drive, C-Partition of Suite PC running out of free space, system remains i" Recalling firm is another firm in the same product code.
first seen 2026-07-08 · recall res_event_number:98096
- recall_reason_pattern
Software/algorithm-related recall in product code OWB (PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Veenpluis 4-6 Best Netherlands, initiated 2025-12-12): "Philips has identified two (2) software issues affecting device systems that may result in loss of imaging (X-ray) functionality and/or loss of motorized movement, and/or incorrect" Recalling firm is another firm in the same product code.
first seen 2026-07-08 · recall res_event_number:98108
- …and 1 more.
Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device). Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.