EPIQ Series Diagnostic Ultrasound System; Affiniti Series Diagnostic Ultrasound System
K250177Philips Ultrasound LLC · cleared 2025-04-10 · product code IYN · Radiology
Premarket evidence — what FDA accepted
source quote (p.6)
“The purpose of this Traditional 510(k) Pre-Market Notification is to introduce Remote Software Management (RSM) software feature onto the EPIQ and Affiniti Series Diagnostic Ultrasound Systems.”
source quote (p.8)
“Remote Software Management provides capabilities to remotely upgrade to major software releases on EPIQ and Affiniti. RSM is capable of deploying the following packages onto the system: OS Packages, Device Drivers, Printer Drivers, Application Packages, Service Application Packages, Automatic backup/restore of existing system settings.”
Validation studies (1)
Bench
sample size not stated
endpoints: All software verification tests met the acceptance criteria.
standards: IEC 62304 Medical Device Software – Software life cycle processes, 2006 + A 2015, ISO 14971 Medical devices - Application of risk management to medical devices
Reported performance (0 observations)
FDA source did not state a quantitative performance metric — non-reporting is itself the signal.
Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.
Predicate network
Postmarket — what happened after clearance
- recall_reason_pattern
Software/algorithm-related recall in product code IYN (Philips Ultrasound, LLC, initiated 2025-10-31): "Ultrasound system compatibility issues with Apple devices running iOS 18 may cause a failure to perform live imagining." Recalling firm matches this device's applicant.
first seen 2026-07-08 · recall res_event_number:97843
- re_clearance
The FDA AI/ML device list shows a newer 510(k) K253595 (decision 2026-03-27) from Philips Ultrasound, LLC for a matching device line ("EPIQ Series Diagnostic Ultrasound System, Affiniti Series Diagnostic Ultrasound System") — a new clearance for the same line is a change event.
first seen 2026-07-08 · k_number:K253595
- re_clearance
The FDA AI/ML device list shows a newer 510(k) K251651 (decision 2025-10-09) from Philips Ultrasound, LLC for a matching device line ("EPIQ Series Diagnostic Ultrasound System; Affiniti Series Diagnostic Ultrasound System") — a new clearance for the same line is a change event.
first seen 2026-07-08 · k_number:K251651
- re_clearance
The FDA AI/ML device list shows a newer 510(k) K251455 (decision 2025-07-24) from Philips Ultrasound LLC for a matching device line ("EPIQ Series Diagnostic Ultrasound System; Affiniti Series Diagnostic Ultrasound System") — a new clearance for the same line is a change event.
first seen 2026-07-08 · k_number:K251455
- re_clearance
The FDA AI/ML device list shows a newer 510(k) K250886 (decision 2025-06-18) from Philips Ultrasound LLC for a matching device line ("EPIQ Series Diagnostic Ultrasound Systems; Affiniti Series Diagnostic Ultrasound Systems") — a new clearance for the same line is a change event.
first seen 2026-07-08 · k_number:K250886
- re_clearance
The FDA AI/ML device list shows a newer 510(k) K243793 (decision 2025-05-21) from Philips Ultrasound LLC for a matching device line ("EPIQ Series Diagnostic Ultrasound System; Affiniti Series Diagnostic Ultrasound System") — a new clearance for the same line is a change event.
first seen 2026-07-08 · k_number:K243793
- …and 4 more.
Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device). Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.
Reimbursement — how devices like this got paid
Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).