uMR 680
K243397Shanghai United Imaging Healthcare Co., Ltd. · cleared 2025-07-16 · product code LNH · Radiology
Premarket evidence — what FDA accepted
source quote (p.6)
“The uMR 680 is a 1.5T superconducting magnetic resonance diagnostic device with a 70cm size patient bore. It consists of components such as magnet, RF power amplifier, RF coils, gradient power amplifier, gradient coils, patient table, spectrometer, computer, equipment cabinets, power distribution system, internal communication system, and vital signal module etc. Addition of imaging reconstruction methods: AI-assisted Compressed Sensing (ACS), Spark artifact Correction (SparkCo).”
source quote (p.14)
“ACS (AI-assisted Compressed Sensing) is an acceleration reconstruction technique. By adding one more regularization term from Al module, ACS is a slight extension of CS (Compressed Sensing). The spark detection module of SparkCo is based on the AI algorithm, however, it won't change the image directly, and it only provides the K-space location of spark points.”
source quote (p.13)
“Content of Premarket Submissions for Management of Cybersecurity in Medical Devices”
Validation studies (6)
Retrospective clinical
n=1,724 images
endpoints: NRMSE; SNR; Resolution (SD/mean value); ROI signal intensities; Image Uniformity; Structure Measurement; Clinical diagnosis quality
standards: NEMA MS 6-2008(R2004)
Retrospective clinical
n=59 cases
endpoints: detection accuracy; PSNR (Peak signal-to-noise ratio) of the spark-corrected images and original spark images; image quality improvement
Retrospective clinical
n=95 cases
endpoints: error between phase indices and gold standard phase indices
Retrospective clinical
n=90 images
endpoints: satisfaction rate S/(S+A+F) exceeding 95%
Retrospective clinical
n=105 cases
endpoints: average Dice coefficient of the left ventricular myocardium after motion correction
Retrospective clinical
n=182 cases
endpoints: average Dice coefficient of the left ventricular myocardium after motion correction
Reported performance (3 observations)
source quote (p.18)
“The average detection accuracy is 94%.”
source quote (p.25)
“The average Dice coefficient of the left ventricular myocardium after motion correction is 0.92, which is greater than 0.87.”
source quote (p.25)
“The average Dice coefficient of the left ventricular myocardium after motion correction is 0.96, which is greater than 0.87.”
Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.
Predicate network
Postmarket — what happened after clearance
- re_clearance
The FDA AI/ML device list shows a newer 510(k) K252371 (decision 2025-09-25) from Shanghai United Imaging Healthcare Co., Ltd. for a matching device line ("uMR 680") — a new clearance for the same line is a change event.
first seen 2026-07-08 · k_number:K252371
- re_clearance
The FDA AI/ML device list shows a newer 510(k) K250246 (decision 2025-08-05) from Shanghai United Imaging Healthcare Co., Ltd. for a matching device line ("uMR Jupiter") — a new clearance for the same line is a change event.
first seen 2026-07-08 · k_number:K250246
- re_clearance
The FDA AI/ML device list shows a newer 510(k) K243547 (decision 2025-07-17) from Shanghai United Imaging Healthcare Co., Ltd. for a matching device line ("uMR Ultra") — a new clearance for the same line is a change event.
first seen 2026-07-08 · k_number:K243547
- recall_reason_pattern
Software/algorithm-related recall in product code LNH (Philips North America, initiated 2026-04-14): "The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction " Recalling firm is another firm in the same product code.
first seen 2026-07-08 · recall res_event_number:98779
- recall_reason_pattern
Software/algorithm-related recall in product code LNH (Philips North America, initiated 2025-12-03): "The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS)." Recalling firm is another firm in the same product code.
first seen 2026-07-08 · recall res_event_number:98111
Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device). Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.
Reimbursement — how devices like this got paid
Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).