Aquilion ONE (TSX-308A/3) V1.5

K242403

Canon Medical Systems Corporation · cleared 2024-12-23 · product code JAK · Radiology

Premarket evidence — what FDA accepted

Device typehardware with ml
source quote (p.7)
Aquilion ONE (TSX-308A/3) V1.5 is a whole body multi-slice helical CT scanner, consisting of a gantry, couch and a console used for data processing and display. AiCE is a noise reduction algorithm that improves image quality and reduces image noise by employing Deep Convolutional Neural Network methods for abdomen, pelvis, lung, cardiac, extremities, head, and inner ear applications.
AlgorithmIterative reconstruction algorithm (FIRST), Deep Convolutional Neural Network methods (AiCE, PIQE, CLEAR Motion), Deep Learning Reconstruction (PIQE, CLEAR Motion)
source quote (p.4)
FIRST is an iterative reconstruction algorithm intended to reduce exposure dose and improve high contrast spatial resolution for abdomen, pelvis, chest, cardiac, extremities and head applications. AiCE is a noise reduction algorithm that improves image quality and reduces image noise by employing Deep Convolutional Neural Network methods for abdomen, pelvis, lung, cardiac, extremities, head, and inner ear applications. PIQE* is a Deep Learning Reconstruction method designed to enhance spatial resolution. By incorporating noise reduction into the Deep Convolutional Neural Network (DCNN), it is possible to achieve both spatial resolution improvement and noise reduction for cardiac, abdomen and pelvis, and lung applications, in comparison to FBP and hybrid iterative reconstruction. CLEAR Motion is a Deep Learning Reconstruction (DLR) method designed to reduce motion artifacts. A Deep Convolutional Neural Network (DCNN) is used to estimate the patient's motion.
Adaptive (vs locked)No
PCCPNo
Cybersecurity addressedYes
source quote (p.9)
Cybersecurity documentation, per the FDA guidance document "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions", issued on September 27, 2023, was included in this submission.

Validation studies (3)

Bench

sample size not stated

endpoints: Contrast-to-Noise Ratios; CT Number Accuracy; Uniformity; Slice Sensitivity Profile; Modulation Transfer Function; Standard Deviation of Noise; Noise Power Spectra; Low Contrast Detectability

Bench

sample size not stated

endpoints: motion artifacts were significantly reduced; CT Numbers were maintained

Retrospective clinical

sample size not stated

endpoints: reconstructed images using the subject device were of diagnostic quality

Reported performance (0 observations)

FDA source did not state a quantitative performance metric — non-reporting is itself the signal.

Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.

Predicate network

Postmarket — what happened after clearance

75
recalls in product code, 24mo
192
MAUDE reports in code, 12mo
-22%
vs code's own 3-yr baseline
9
drift signals on this device
  • recall_reason_pattern

    Software/algorithm-related recall in product code JAK (GE Medical Systems, LLC, initiated 2026-03-26): "GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with " Recalling firm matches this device's applicant.

    first seen 2026-07-08 · recall res_event_number:98738

  • recall_reason_pattern

    Software/algorithm-related recall in product code JAK (PHILIPS MEDICAL SYSTEMS, initiated 2026-03-07): "Philips has identified three software issues: 1. During a continuous CT (CCT) scan, there is the potential that the Gantry could remain at the current scan position after pressing" Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:98588

  • recall_reason_pattern

    Software/algorithm-related recall in product code JAK (Siemens Medical Solutions USA, Inc, initiated 2025-12-19): "To remove the software applications from certain CT systems as the applications have not received FDA 510(k) clearance." Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:98206

  • recall_reason_pattern

    Software/algorithm-related recall in product code JAK (PHILIPS MEDICAL SYSTEMS, initiated 2025-09-25): "Issue 1: The potential for unintentional continued gantry/couch movement when a specific button series is used requiring use of manual stop. Issue 2. When performing a helical/Axia" Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:97699

  • recall_reason_pattern

    Software/algorithm-related recall in product code JAK (Philips North America Llc, initiated 2025-05-29): "Devices with affected software may experience two unintended motion issues that may lead to contact between the Gantry or table with the operator or patient, along with additional " Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:97010

  • recall_reason_pattern

    Software/algorithm-related recall in product code JAK (, initiated 2025-05-23): "Due to a software bug the robotic arm may incorrectly guide the instrument compared to the planning defined by user." Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:96967

  • …and 3 more.

Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device). Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.

Reimbursement — how devices like this got paid

Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).