6450 Ultrasound System (MyLabE80); 6450 Ultrasound System (MyLabE85)

K241671

Esaote S.p.A. · cleared 2025-05-16 · product code IYN · Radiology

Premarket evidence — what FDA accepted

Device typehardware
source quote (p.6)
6450 Ultrasound System is a general-purpose diagnostic ultrasound system, based on a mainframe platform that can be easily moved thanks to four swivelling wheels.
AlgorithmXStrain LA is an advanced processing package that allows clinicians to quantify endocardial velocities of contraction and relaxation and local deformation of the heart (Strain/Strain rate) for Left Atrium analysis, with manual segmentation.
source quote (p.7)
XStrain LA: XStrain allows clinicians to quantify endocardial velocities of contraction and relaxation and local deformation of the heart (Strain/Strain rate). XStrain LA is an advanced processing package for the Left Atrium analysis. Strain functionality Left atrium (LA) with manual segmentation.
Adaptive (vs locked)FDA source did not state this
PCCPFDA source did not state this
Cybersecurity addressedFDA source did not state this

Validation studies (2)

Bench

sample size not stated

endpoints: acoustic output; biocompatibility; cleaning and disinfection effectiveness; thermal safety; electrical safety; electromagnetic safety; mechanical safety

standards: ANSI/AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012, IEC 60601-1-2:2014 + AMD1:2020 (Ed. 4.1), IEC 60601-1-6:2010+A1:2013, IEC 60601-2-37:2007 +A1:2015, NEMA UD 2-2004 (R2009), NEMA UD 3-2004 (R2009)

Retrospective clinical

sample size not stated · 2 site(s)

endpoints: Comparison of LASr, LAScd, and LASct results to TomTec reference; Intraclass Correlation Coefficient (ICC) for LASr, LAScd, and LASct

Reported performance (3 observations)

agreement_kappaas written: “ICC for LASr0.84
source quote (p.8)
The ICC, Intraclass Correlation Coefficient result is valid for all Left Atrium Strain indexes (LASr, LAScd and LASct): LASr = 0.84
agreement_kappaas written: “ICC for LAScd0.76
source quote (p.8)
The ICC, Intraclass Correlation Coefficient result is valid for all Left Atrium Strain indexes (LASr, LAScd and LASct): LAScd = 0,76
agreement_kappaas written: “ICC for LASct0.74
source quote (p.8)
The ICC, Intraclass Correlation Coefficient result is valid for all Left Atrium Strain indexes (LASr, LAScd and LASct): LASct = 0,74

Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.

Predicate network

Postmarket — what happened after clearance

85
recalls in product code, 24mo
554
MAUDE reports in code, 12mo
+85%
vs code's own 3-yr baseline
4
drift signals on this device
  • recall_reason_pattern

    Software/algorithm-related recall in product code IYN (Philips Ultrasound, LLC, initiated 2025-10-31): "Ultrasound system compatibility issues with Apple devices running iOS 18 may cause a failure to perform live imagining." Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:97843

  • recall_reason_pattern

    Software/algorithm-related recall in product code IYN (GE Medical Systems, LLC, initiated 2025-09-18): "The Ultrasound-Guided Attenuation Parameter (UGAP) measurement data may display inaccurate values representing liver steatosis. This could potentially lead to inappropriate clinica" Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:97726

  • recall_reason_pattern

    Software/algorithm-related recall in product code IYN (GE Medical Systems China Co., Ltd. Dev. Zone National Hi-Tech; No., initiated 2025-05-16): "GE HealthCare has become aware that the Estimated Fetal Weight (EFW) measurement data feature on the Versana Premier R3 and LOGIQ F R3 series ultrasound systems can display previou" Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:96992

  • recall_reason_pattern

    Software/algorithm-related recall in product code IYN (Siemens Medical Solutions USA, Inc., initiated 2024-08-15): "If ultrasound systems with software, are changed from factory default to : 1) Milliliters per second (ml/sec, mL/sec) or 2) Milliliters per minute (ml/min, mL/min); then systems wi" Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:95254

Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device). Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.

Reimbursement — how devices like this got paid

Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).