MuscleView
K241331Springbok, Inc. · cleared 2024-10-01 · product code LNH · Radiology
Premarket evidence — what FDA accepted
source quote (p.6)
“MuscleView is a software only product that uses a machine learning-based approach for the automatic segmentation of musculoskeletal structures from MRI.”
source quote (p.6)
“MuscleView is a software only product that uses a machine learning-based approach for the automatic segmentation of musculoskeletal structures from MRI. ... A library of already contoured expert cases is utilized to train the machine learning algorithms, specifically convolutional neural networks (CNNs) perform automated segmentation.”
source quote (p.6)
“The segmentation & 3D model viewer module handles training data and algorithms to obtain the pre-trained models and algorithms to update models.”
Validation studies (1)
Retrospective clinical
n=148 cases · 19 site(s)
endpoints: dice similarity coefficient (DSC) below interobserver variability; volume difference (VDt) below interobserver variability
Reported performance (3 observations)
source quote (p.13)
“adductor brevis 0.962 -0.965”
source quote (p.13)
“gluteus maximus 0.993-0.994”
source quote (p.13)
“pelvis 0.978-0.981”
Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.
Predicate network
Postmarket — what happened after clearance
- recall_reason_pattern
Software/algorithm-related recall in product code LNH (Philips North America, initiated 2026-04-14): "The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction " Recalling firm is another firm in the same product code.
first seen 2026-07-08 · recall res_event_number:98779
- recall_reason_pattern
Software/algorithm-related recall in product code LNH (Philips North America, initiated 2025-12-03): "The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS)." Recalling firm is another firm in the same product code.
first seen 2026-07-08 · recall res_event_number:98111
Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device). Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.
Reimbursement — how devices like this got paid
Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).