CINA-VCF
K240612Avicenna.AI · cleared 2024-05-31 · product code QFM · Radiology
Premarket evidence — what FDA accepted
source quote (p.3)
“CINA-VCF is a radiological computer aided triage and notification software indicated for use in patients aged 50 years and over undergoing non-enhanced or contrast-enhanced CT scans which include the chest and/or abdomen.”
source quote (p.5)
“The device uses deep learning models to detect VCF at the T1-L5 level.”
source quote (p.10)
“CINA-VCF has similar technological characteristics as the predicate device, i.e., the deployment of an artificial intelligence algorithm with a database of images.”
Validation studies (1)
Retrospective clinical
n=474 cases
endpoints: evaluate the software's performance in identifying vertebral compression fractures (VCF); CINA-VCF time-to-notification, which was compared to the predicate device for true positive cases only
standards: NEMA PS 3.1 - 3.20
Reported performance (7 observations)
source quote (p.7)
“Sensitivity and Specificity were 95.2% [95% CI: 90.7% – 97.9%]”
source quote (p.7)
“and 92.9% [95% CI: 89.4% – 96.5%], respectively.”
source quote (p.7)
“The ROC AUC was 0.974 [95% CI: 0.962 - 0.986], which exceeded the 0.95 performance goal, thus, achieving the primary endpoint.”
source quote (p.7)
“Similarly, the overall agreement (Accuracy) was 93.7% [95% CI: 91.1% - 95%], which represents very good predictions.”
source quote (p.7)
“The mean [95% Cl] time-to-notification for all included cases (n = 474) was estimated to be 23.4 [95% CI: 22.7 – 24.2] seconds for CINA-VCF.”
source quote (p.8)
“When taking into account only true positive cases (n = 158), the mean [95% CI] time-to-notification was 21.7 [95% CI: 20.5 – 22.9] seconds for CINA-VCF”
source quote (p.8)
“and 117.2 [95% CI: 98.64 – 135.85] seconds for BriefCase, the selected predicate device.”
Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.
Predicate network
Postmarket — what happened after clearance
Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device). Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.
Reimbursement — how devices like this got paid
Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).