SOMATOM go.Up; SOMATOM go.Now; SOMATOM go.All; SOMATOM go.Top; SOMATOM go.Sim; SOMATOM go.Open Pro; SOMATOM X.cite; SOMATOM X.ceed
K233650Siemens Medical Solutions USA, Inc. · cleared 2024-03-26 · product code JAK · Radiology
Premarket evidence — what FDA accepted
source quote (p.6)
“The subject devices SOMATOM go. Platform and SOMATOM X. Platform with SOMARIS/10 syngo CT VB10 are Computed Tomography X-ray Systems which feature one continuously rotating tube-detector system and function according to the fan beam principle (single source).”
source quote (p.9)
“Iterative Reconstruction – Support of various iterative reconstruction principles”
source quote (p.39)
“Siemens conforms to the Cybersecurity requirements by implementing a process of preventing unauthorized access, modifications, misuse or denial of use, or the unauthorized use of information that is stored, accessed, or transferred from a medical device to an external recipient. Cybersecurity information in accordance with guidance document "Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions" issued on September 27, 2023 is included within this submission.”
Validation studies (1)
Bench
sample size not stated
endpoints: verify and validate the functionality of the subject device modifications; test specification and acceptance criteria are related to the corresponding requirements
standards: NEMA PS 3.1 - 3.20 2022d, NEMA XR 25-2019, NEMA XR 28-2018, IEC 61223-3-5 Edition 2.0 2019-09, IEC 61223-2-6 Second Edition 2006-11, IEC 60601-1-3 Edition 2.1 2013-04, IEC 60601-2-44 Edition 3.2: 2016, ANSI AAMI ISO 14971: 2019, ISO 14971 Third Edition 2019-12, ANSI AAMI IEC 62304:2006/A1:2016, IEC 62304 Edition 1.1 2015-06, ANSI AAMI ES60601-1:2005/(R)2012 & A1:2012, C1:2009/(R)2012 & A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021], ANSI AAMI IEC 60601-1-2:2014 [Including AMD 1:2021], IEC 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION, ANSI AAMI IEC 62366-1:2015+AMD1:2020 (Consolidated Text), IEC 62366-1 Edition 1.1 2020-06 CONSOLIDATED VERSION, IEC 60825-1 Edition 2.0 2007-03, IEC 60601-1-6 Edition 3.2 2020-07 CONSOLIDATED VERSION, IEC 60601-2-28 Edition 3.0 2017-06, IEC 62563-1 Edition 1.2 2021-07 CONSOLIDATED VERSION, IEC 60601-1:2005+A1:2012+A2:2020, IEC/ISO 17050-1, IEC/ISO 17050-2
Reported performance (1 observation)
source quote (p.34)
“It has been shown that the algorithm was able to consistently detect needle-tips over a wide variety of scans in 90.76% of cases.”
Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.
Predicate network
Postmarket — what happened after clearance
- recall_reason_pattern
Software/algorithm-related recall in product code JAK (Siemens Medical Solutions USA, Inc, initiated 2025-12-19): "To remove the software applications from certain CT systems as the applications have not received FDA 510(k) clearance." Recalling firm matches this device's applicant.
first seen 2026-07-08 · recall res_event_number:98206
- re_clearance
The FDA AI/ML device list shows a newer 510(k) K253574 (decision 2026-03-20) from Siemens Medical Solutions USA, Inc. for a matching device line ("SOMATOM X.cite; SOMATOM X.ceed") — a new clearance for the same line is a change event.
first seen 2026-07-08 · k_number:K253574
- recall_reason_pattern
Software/algorithm-related recall in product code JAK (GE Medical Systems, LLC, initiated 2026-03-26): "GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with " Recalling firm is another firm in the same product code.
first seen 2026-07-08 · recall res_event_number:98738
- recall_reason_pattern
Software/algorithm-related recall in product code JAK (PHILIPS MEDICAL SYSTEMS, initiated 2026-03-07): "Philips has identified three software issues: 1. During a continuous CT (CCT) scan, there is the potential that the Gantry could remain at the current scan position after pressing" Recalling firm is another firm in the same product code.
first seen 2026-07-08 · recall res_event_number:98588
- recall_reason_pattern
Software/algorithm-related recall in product code JAK (PHILIPS MEDICAL SYSTEMS, initiated 2025-09-25): "Issue 1: The potential for unintentional continued gantry/couch movement when a specific button series is used requiring use of manual stop. Issue 2. When performing a helical/Axia" Recalling firm is another firm in the same product code.
first seen 2026-07-08 · recall res_event_number:97699
- recall_reason_pattern
Software/algorithm-related recall in product code JAK (Philips North America Llc, initiated 2025-05-29): "Devices with affected software may experience two unintended motion issues that may lead to contact between the Gantry or table with the operator or patient, along with additional " Recalling firm is another firm in the same product code.
first seen 2026-07-08 · recall res_event_number:97010
- …and 4 more.
Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device). Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.