CINA-ASPECTS
K233342Avicenna.AI · cleared 2024-03-15 · product code POK · Radiology
Premarket evidence — what FDA accepted
source quote (p.3)
“CINA-ASPECTS is a computer-aided diagnosis (CADx) software device used to assist the clinician in the assessment and characterization of brain tissue abnormalities using CT image data.”
source quote (p.9)
“CINA-ASPECTS uses machine learning algorithms based on deep learning techniques.”
Validation studies (3)
Bench
sample size not stated
standards: NEMA PS 3.1 - 3.20
Retrospective clinical
n=200 cases · 5 site(s)
endpoints: generalize to a range of typical patient demographics, Clinical parameters, ASPECTS regions, and image acquisition parameters
Reader study (MRMC)
n=200 cases
endpoints: assessment of ASPECTS; improving reader accuracy
Reported performance (1 observation)
source quote (p.8)
“In addition, the overall readers' ROC AUC also improved significantly from 0.75 (Unaided arm) to 0.79 (Aided arm).”
Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.
Predicate network
Postmarket — what happened after clearance
Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device). Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.
Reimbursement — how devices like this got paid
Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).