CT Cardiomegaly

K232613

Innolitics, LLC · cleared 2024-02-28 · product code QIH · Radiology

Premarket evidence — what FDA accepted

Device typesamd
source quote (p.5)
The CT Cardiomegaly application is a software only (SaMD) medical device which includes automated algorithms and non-adaptive machine learning to analyze chest computed tomography (CT) images.
AlgorithmMachine learning based algorithm
source quote (p.11)
Machine learning based algorithm (non-adaptive)
Adaptive (vs locked)No
source quote (p.6)
CT Cardiomegaly is designed to measure the maximal transverse diameter of heart and maximal inner transverse diameter of thoracic cavity and calculate the CTR from an axial CT slice containing the heart using a non-adaptive machine learning algorithm.
PCCPFDA source did not state this
Cybersecurity addressedFDA source did not state this

Validation studies (1)

Retrospective clinical

n=275 cases

endpoints: calculating the difference between the average measurement of the human readers and the subject device's measurement; precision of selecting the slice where the heart area is the largest (key heart slice detection); accuracy of heart and inner chest segmentations

Reported performance (2 observations)

diceas written: “Heart Observed Dice0.95CI [0.950, 0.956]
source quote (p.13)
0.95 [0.950, 0.956]
diceas written: “Inner Chest Observed Dice0.98CI [0.982, 0.984]
source quote (p.13)
0.98 [0.982, 0.984]

Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.

Predicate network

Postmarket — what happened after clearance

0
recalls in product code, 24mo
3
MAUDE reports in code, 12mo
vs code's own 3-yr baseline
0
drift signals on this device

Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device). Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.

Reimbursement — how devices like this got paid

Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).

RIGOR™ Precedent · public FDA/CMS data · descriptive decision-support, not regulatory or reimbursement advice. Share this page: radar.healthai.com/precedent/device/K232613