MAGNETOM Avanto fit; MAGNETOM Skyra fit
K232494Siemens Medical Solutions USA, Inc. · cleared 2023-11-14 · product code LNH · Radiology
Premarket evidence — what FDA accepted
source quote (p.4)
“The MAGNETOM system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities.”
Validation studies (1)
Bench
sample size not stated
endpoints: Software verification and validation; Sample clinical images; Image quality assessment by sample clinical images; SNR and image uniformity measurements for coils; Heating measurements for coils
standards: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, Guidance for submission of Premarket Notifications for Magnetic Resonance Diagnostic Devices
Reported performance (0 observations)
FDA source did not state a quantitative performance metric — non-reporting is itself the signal.
Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.
Predicate network
Postmarket — what happened after clearance
- recall_reason_pattern
Software/algorithm-related recall in product code LNH (Philips North America, initiated 2026-04-14): "The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction " Recalling firm is another firm in the same product code.
first seen 2026-07-08 · recall res_event_number:98779
- recall_reason_pattern
Software/algorithm-related recall in product code LNH (Philips North America, initiated 2025-12-03): "The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS)." Recalling firm is another firm in the same product code.
first seen 2026-07-08 · recall res_event_number:98111
Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device). Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.