LOGIQ Totus

K232381

GE Medical Systems Ultrasound and Primary care Diagnostics, · cleared 2023-12-07 · product code IYN · Radiology

Premarket evidence — what FDA accepted

Device typehardware with ml
source quote (p.4)
The LOGIQ Totus is full featured, Track 3 device, primarily intended for general purpose diagnostic ultrasound system which consists of a mobile console approximately 490mm wide(monitor width: 545mm), 835mm deep and 1415~1815mm high that provides digital acquisition, processing and display capability. AI Testing Summary: Auto preset selection: Auto Abdominal Color Assistant:
AlgorithmAuto preset selection and Auto Abdominal Color Assistant
source quote (p.7)
AI Testing Summary: Auto preset selection: Auto Abdominal Color Assistant:
Adaptive (vs locked)FDA source did not state this
PCCPFDA source did not state this
Cybersecurity addressedFDA source did not state this

Validation studies (2)

Retrospective clinical

n=50 patients

endpoints: Overall model success rate of view suggestion

Retrospective clinical

n=40 patients

endpoints: Overall model success rate of view suggestion

Reported performance (0 observations)

FDA source did not state a quantitative performance metric — non-reporting is itself the signal.

Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.

Predicate network

Postmarket — what happened after clearance

85
recalls in product code, 24mo
554
MAUDE reports in code, 12mo
+85%
vs code's own 3-yr baseline
5
drift signals on this device
  • recall_reason_pattern

    Software/algorithm-related recall in product code IYN (GE Medical Systems, LLC, initiated 2025-09-18): "The Ultrasound-Guided Attenuation Parameter (UGAP) measurement data may display inaccurate values representing liver steatosis. This could potentially lead to inappropriate clinica" Recalling firm matches this device's applicant.

    first seen 2026-07-08 · recall res_event_number:97726

  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K253370 (decision 2026-01-08) from GE Medical Systems Ultrasound and Primary Care Diagnostics for a matching device line ("LOGIQ Totus") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K253370

  • recall_reason_pattern

    Software/algorithm-related recall in product code IYN (Philips Ultrasound, LLC, initiated 2025-10-31): "Ultrasound system compatibility issues with Apple devices running iOS 18 may cause a failure to perform live imagining." Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:97843

  • recall_reason_pattern

    Software/algorithm-related recall in product code IYN (GE Medical Systems China Co., Ltd. Dev. Zone National Hi-Tech; No., initiated 2025-05-16): "GE HealthCare has become aware that the Estimated Fetal Weight (EFW) measurement data feature on the Versana Premier R3 and LOGIQ F R3 series ultrasound systems can display previou" Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:96992

  • recall_reason_pattern

    Software/algorithm-related recall in product code IYN (Siemens Medical Solutions USA, Inc., initiated 2024-08-15): "If ultrasound systems with software, are changed from factory default to : 1) Milliliters per second (ml/sec, mL/sec) or 2) Milliliters per minute (ml/min, mL/min); then systems wi" Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:95254

Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device). Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.

Reimbursement — how devices like this got paid

Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).