Contour ProtégéAI
K231765MIM Software Inc. · cleared 2023-11-08 · product code QKB · Radiology
Premarket evidence — what FDA accepted
source quote (p.4)
“Contour ProtégéAI is an accessory to MIM software that automatically creates contours on medical images through the use of machine-learning based algorithms.”
source quote (p.4)
“Contour ProtégéAI is an accessory to MIM software that automatically creates contours on medical images through the use of machine-learning algorithms. A total of 550 CT images from 41 clinical sites across multiple continents was gathered for the training of the final 4.1.0 neural network models.”
source quote (p.5)
“A total of 550 CT images from 41 clinical sites across multiple continents was gathered for the training of the final 4.1.0 neural network models.”
Validation studies (1)
Retrospective clinical
n=754 images · 27 site(s)
endpoints: Dice score; MDA score; Average user evaluation (time savings); Cumulative Added Path Length; localization accuracy
Reported performance (2 observations)
source quote (p.13)
“0.86 ± 0.07 (0.83) *”
source quote (p.13)
“0.22 ± 0.10 (0.14) *”
Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.
Predicate network
Postmarket — what happened after clearance
- re_clearance
The FDA AI/ML device list shows a newer 510(k) K253270 (decision 2026-03-27) from Mim Software, Inc. for a matching device line ("Contour ProtégéAI+") — a new clearance for the same line is a change event. The newer clearance's parsed summary mentions a PCCP.
first seen 2026-07-08 · k_number:K253270
- re_clearance
The FDA AI/ML device list shows a newer 510(k) K250035 (decision 2025-02-03) from MIM Software Inc. for a matching device line ("Contour ProtégéAI+") — a new clearance for the same line is a change event.
first seen 2026-07-08 · k_number:K250035
Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device). Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.
Reimbursement — how devices like this got paid
Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).