Lunit INSIGHT DBT

K231470

Lunit Inc. · cleared 2023-11-06 · product code QDQ · Radiology

Premarket evidence — what FDA accepted

Device typesamd
source quote (p.5)
Lunit INSIGHT DBT is a computer-assisted detection and diagnosis (CADe/x) software as a medical device that provides information about the presence, location and characteristics of lesions suspicious for breast cancer to assist interpreting physicians in making diagnostic decisions when reading digital breast tomosynthesis (DBT) images. The software automatically analyzes digital breast tomosynthesis slices via artificial intelligence technology that has been trained via deep learning.
Algorithmartificial intelligence technology that has been trained via deep learning
source quote (p.5)
The software automatically analyzes digital breast tomosynthesis slices via artificial intelligence technology that has been trained via deep learning.
Adaptive (vs locked)FDA source did not state this
PCCPFDA source did not state this
Cybersecurity addressedNo

Validation studies (2)

Standalone

n=2,202 cases

endpoints: demonstrate AUROC in standalone performance greater than 0.903

Reader study (MRMC)

n=258 cases

endpoints: evaluate the effectiveness of Lunit INSIGHT DBT by comparing the clinical performance of radiologists with CAD and without CAD assistance

Reported performance (3 observations)

aurocas written: “auc0.928CI 0.917 - 0.939
source quote (p.8)
The subject device's AUROC in the standalone performance analysis was 0.928 (95% CI: 0.917 - 0.939) with statistical significance (p < 0.0001), which exceeded the acceptance criteria of the primary endpoint.
aurocas written: “AUROC for CAD-unassisted interpretation0.897CI 0.858 - 0.936
source quote (p.8)
AUROC for CAD-unassisted interpretation was 0.897 (95% CI 0.858 - 0.936)
aurocas written: “AUROC for CAD-assisted interpretation0.915CI 0.874 - 0.955
source quote (p.8)
when that of CAD-assisted interpretation was 0.915 (95% CI: 0.874 - 0.955)

Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.

Predicate network

Postmarket — what happened after clearance

0
recalls in product code, 24mo
0
MAUDE reports in code, 12mo
vs code's own 3-yr baseline
2
drift signals on this device
  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K253796 (decision 2026-03-26) from Lunit, Inc. for a matching device line ("Lunit INSIGHT DBT (V1.2)") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K253796

  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K242652 (decision 2024-10-04) from Lunit Inc. for a matching device line ("Lunit INSIGHT DBT v1.1") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K242652