Lunit INSIGHT DBT
K231470Lunit Inc. · cleared 2023-11-06 · product code QDQ · Radiology
Premarket evidence — what FDA accepted
source quote (p.5)
“Lunit INSIGHT DBT is a computer-assisted detection and diagnosis (CADe/x) software as a medical device that provides information about the presence, location and characteristics of lesions suspicious for breast cancer to assist interpreting physicians in making diagnostic decisions when reading digital breast tomosynthesis (DBT) images. The software automatically analyzes digital breast tomosynthesis slices via artificial intelligence technology that has been trained via deep learning.”
source quote (p.5)
“The software automatically analyzes digital breast tomosynthesis slices via artificial intelligence technology that has been trained via deep learning.”
Validation studies (2)
Standalone
n=2,202 cases
endpoints: demonstrate AUROC in standalone performance greater than 0.903
Reader study (MRMC)
n=258 cases
endpoints: evaluate the effectiveness of Lunit INSIGHT DBT by comparing the clinical performance of radiologists with CAD and without CAD assistance
Reported performance (3 observations)
source quote (p.8)
“The subject device's AUROC in the standalone performance analysis was 0.928 (95% CI: 0.917 - 0.939) with statistical significance (p < 0.0001), which exceeded the acceptance criteria of the primary endpoint.”
source quote (p.8)
“AUROC for CAD-unassisted interpretation was 0.897 (95% CI 0.858 - 0.936)”
source quote (p.8)
“when that of CAD-assisted interpretation was 0.915 (95% CI: 0.874 - 0.955)”
Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.
Predicate network
Postmarket — what happened after clearance
- re_clearance
The FDA AI/ML device list shows a newer 510(k) K253796 (decision 2026-03-26) from Lunit, Inc. for a matching device line ("Lunit INSIGHT DBT (V1.2)") — a new clearance for the same line is a change event.
first seen 2026-07-08 · k_number:K253796
- re_clearance
The FDA AI/ML device list shows a newer 510(k) K242652 (decision 2024-10-04) from Lunit Inc. for a matching device line ("Lunit INSIGHT DBT v1.1") — a new clearance for the same line is a change event.
first seen 2026-07-08 · k_number:K242652
Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device). Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.
Reimbursement — how devices like this got paid
Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).