HepaFatSmart (V2.0.0)

K231459

Resonance Health Analysis Services Pty Ltd · cleared 2023-06-20 · product code LNH · Radiology

Premarket evidence — what FDA accepted

Device typesamd
source quote (p.5)
HepaFatSmart is an SaMD designed to automatically analyse magnetic resonance imaging (MRI) datasets for quantitative assessment of a patient's liver fat, in form of volumetric liver fat fraction (VLFF), proton density fat fraction (PDFF), and steatosis grade. It is an AI assisted, automated version of HepaFat-Scan (another SaMD of Resonance Health).
Algorithmconvolutional neural networks
source quote (p.5)
It is composed of one (1) convolutional neural network (CNN) performing liver ROI detection with undesired components (artefacts and major blood vessels) considered/removed using a computer vision technique with machine learning technology.
Adaptive (vs locked)No
source quote (p.5)
Following the training of the AI assisted device, the system is completely 'locked down' for final validation prior to release in commercial use to ensure reproducibility of the results.
PCCPFDA source did not state this
Cybersecurity addressedFDA source did not state this

Validation studies (2)

Retrospective clinical

n=41 patients

endpoints: VLFF

Retrospective clinical

n=281 patients

endpoints: VLFF; Sensitivity; Specificity

Reported performance (2 observations)

sensitivity100CI 97.3 - 100.0
source quote (p.14)
100.0 (97.3 - 100.0)
specificity98.6CI 94.9 - 99.6
source quote (p.14)
98.6 (94.9 - 99.6)

Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.

Predicate network

Postmarket — what happened after clearance

100
recalls in product code, 24mo
510
MAUDE reports in code, 12mo
+5%
vs code's own 3-yr baseline
2
drift signals on this device
  • recall_reason_pattern

    Software/algorithm-related recall in product code LNH (Philips North America, initiated 2026-04-14): "The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction " Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:98779

  • recall_reason_pattern

    Software/algorithm-related recall in product code LNH (Philips North America, initiated 2025-12-03): "The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS)." Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:98111

Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device). Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.

Reimbursement — how devices like this got paid

Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).

RIGOR™ Precedent · public FDA/CMS data · descriptive decision-support, not regulatory or reimbursement advice. Share this page: radar.healthai.com/precedent/device/K231459