Rapid Aneurysm Triage and Notification

K230074

iSchemaView Inc. · cleared 2023-07-27 · product code QFM · Radiology

Premarket evidence — what FDA accepted

Device typesamd
source quote (p.3)
Rapid Aneurysm Triage and Notification (ANRTN) is a radiological computer-assisted triage and notification software device for analysis of CT images of the head.
Algorithmartificial intelligence algorithm
source quote (p.3)
Rapid ANRTN uses an artificial intelligence algorithm to analyze images and highlight studies with suspected saccular aneurysms in a standalone application for study list prioritization or triage in parallel to ongoing standard of care.
Adaptive (vs locked)FDA source did not state this
PCCPFDA source did not state this
Cybersecurity addressedYes
source quote (p.11)
Rapid ANRTN has been developed using the cybersecurity framework defined within the FDA Draft Guidance: Cybersecurity in Medical Devices: Quality System Considerations and Content of Premarket Submissions

Validation studies (1)

Standalone

n=266 cases

endpoints: AUC ">= 0.95 for high performance

standards: EN ISO 14971:2019 (R2021), IEC 62304:2006 (R2015), IEC 62366:2015 (R2020), NEMA PS 3.1 - 3.20

Reported performance (2 observations)

sensitivity0.933
source quote (p.9)
Additionally, Sensitivity (0.933) and Specificity(0.868) supported the finding.
specificity0.868
source quote (p.9)
Additionally, Sensitivity (0.933) and Specificity(0.868) supported the finding.

Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.

Predicate network

Postmarket — what happened after clearance

0
recalls in product code, 24mo
0
MAUDE reports in code, 12mo
vs code's own 3-yr baseline
1
drift signals on this device
  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K233512 (decision 2024-01-16) from iSchemaView, Inc. for a matching device line ("Rapid (6.0)") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K233512

Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device). Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.

Reimbursement — how devices like this got paid

Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).