DeepXray

K223621

Alpha Intelligence Manifolds, Inc. · cleared 2023-09-08 · product code QIH · Radiology

Premarket evidence — what FDA accepted

Device typesamd
source quote (p.3)
DeepXray is a radiological fully automated image processing software device of either computed (CR) or directly digital (DX) images intended to aid medical professionals in the measurement of minimum joint space width; the assessment of the presence or absence of sclerosis, joint space narrowing, and osteophytes based on OARSI criteria for these parameters; and, the presence or absence of radiographic knee OA based on Kellgren & Lawrence Grading of standing, fixed-flexion radiographs of the knee.
Algorithmartificial intelligence (AI) and computer vision algorithms
source quote (p.4)
DeepXray is a standalone software device that utilizes artificial intelligence (AI) and computer vision algorithms to assist clinical professionals in analyzing and measuring radiographic abnormalities of knee osteoarthritis (OA) during review of posterior-anterior or anterior-posterior knee radiographs.
Adaptive (vs locked)FDA source did not state this
PCCPFDA source did not state this
Cybersecurity addressedFDA source did not state this

Validation studies (1)

Retrospective clinical

n=1,121 patients · 1 site(s)

endpoints: Kellgren-Lawrence Grade assessment; Joint Space Narrowing assessment; Osteophyte assessment; Sclerosis assessment; Medial mJSW measurement; Lateral mJSW measurement; Femoral-Tibial Angle measurement

Reported performance (8 observations)

sensitivity0.87CI 0.86/0.88
source quote (p.8)
Sensitivity (KL Grade ≥2) 0.87 (0.86/0.88)
specificity0.84CI 0.83/0.85
source quote (p.8)
Specificity (KL Grade ≥2) 0.84 (0.83/0.85)
sensitivityas written: “Sensitivity (OARSI Grade ≥1) for Joint Space Narrowing0.88CI 0.87/0.89
source quote (p.8)
Sensitivity (OARSI Grade ≥1) 0.88 (0.87/0.89)
specificityas written: “Specificity (OARSI Grade ≥1) for Joint Space Narrowing0.82CI 0.81/0.83
source quote (p.8)
Specificity (OARSI Grade ≥1) 0.82 (0.81/0.83)
sensitivityas written: “Sensitivity (OARSI Grade ≥1) for Osteophyte0.86CI 0.85/0.87
source quote (p.8)
Sensitivity (OARSI Grade ≥1) 0.86 (0.85/0.87)
specificityas written: “Specificity (OARSI Grade ≥1) for Osteophyte0.8CI 0.79/0.81
source quote (p.8)
Specificity (OARSI Grade ≥1) 0.80 (0.79/0.81)
sensitivityas written: “Sensitivity (Presence/Absence) for Sclerosis0.84CI 0.83/0.85
source quote (p.8)
Sensitivity (Presence/Absence) 0.84 (0.83/0.85)
specificityas written: “Specificity (Presence/Absence) for Sclerosis0.88CI 0.87/0.89
source quote (p.8)
Specificity (Presence/Absence) 0.88 (0.87/0.89)

Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.

Predicate network

Postmarket — what happened after clearance

0
recalls in product code, 24mo
3
MAUDE reports in code, 12mo
vs code's own 3-yr baseline
0
drift signals on this device

Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device). Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.

Reimbursement — how devices like this got paid

Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).

RIGOR™ Precedent · public FDA/CMS data · descriptive decision-support, not regulatory or reimbursement advice. Share this page: radar.healthai.com/precedent/device/K223621