Spectral Bone Marrow
K223514GE Medical Systems, LLC · cleared 2023-03-09 · product code JAK · Radiology
Premarket evidence — what FDA accepted
source quote (p.3)
“Spectral Bone Marrow is an automated image processing software application, utilizing deep learning technology for bone segmentation, to facilitate optimized visualization of bone marrow in spectral body and extremity CT images.”
source quote (p.4)
“Spectral Bone Marrow is a deep-learning based software analysis package designed for the visualization of bone marrow based on GE HealthCare's spectral CT acquisitions data.”
Validation studies (2)
Bench
n=146 cases
endpoints: algorithm ability to accurately segment bone
Reader study (MRMC)
sample size not stated
endpoints: assessment of image quality related to diagnostic use according to a Likert scale; rate their efficiency when using the algorithm compared to using without
Reported performance (0 observations)
FDA source did not state a quantitative performance metric — non-reporting is itself the signal.
Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.
Predicate network
Postmarket — what happened after clearance
- recall_reason_pattern
Software/algorithm-related recall in product code JAK (GE Medical Systems, LLC, initiated 2026-03-26): "GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with " Recalling firm matches this device's applicant.
first seen 2026-07-08 · recall res_event_number:98738
- recall_reason_pattern
Software/algorithm-related recall in product code JAK (PHILIPS MEDICAL SYSTEMS, initiated 2026-03-07): "Philips has identified three software issues: 1. During a continuous CT (CCT) scan, there is the potential that the Gantry could remain at the current scan position after pressing" Recalling firm matches this device's applicant.
first seen 2026-07-08 · recall res_event_number:98588
- recall_reason_pattern
Software/algorithm-related recall in product code JAK (PHILIPS MEDICAL SYSTEMS, initiated 2025-09-25): "Issue 1: The potential for unintentional continued gantry/couch movement when a specific button series is used requiring use of manual stop. Issue 2. When performing a helical/Axia" Recalling firm matches this device's applicant.
first seen 2026-07-08 · recall res_event_number:97699
- recall_reason_pattern
Software/algorithm-related recall in product code JAK (Philips Medical Systems Nederland B.V. Veenpluis, initiated 2025-02-28): "Multiple problems identified with the software version leading to various scanning and image issues, and unintended device movement." Recalling firm matches this device's applicant.
first seen 2026-07-08 · recall res_event_number:96423
- recall_reason_pattern
Software/algorithm-related recall in product code JAK (Siemens Medical Solutions USA, Inc, initiated 2025-12-19): "To remove the software applications from certain CT systems as the applications have not received FDA 510(k) clearance." Recalling firm is another firm in the same product code.
first seen 2026-07-08 · recall res_event_number:98206
- recall_reason_pattern
Software/algorithm-related recall in product code JAK (Philips North America Llc, initiated 2025-05-29): "Devices with affected software may experience two unintended motion issues that may lead to contact between the Gantry or table with the operator or patient, along with additional " Recalling firm is another firm in the same product code.
first seen 2026-07-08 · recall res_event_number:97010
- …and 3 more.
Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device). Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.
Reimbursement — how devices like this got paid