ECHELON Synergy MRI system
K223426FUJIFILM Healthcare Corporation · cleared 2023-07-13 · product code LNH · Radiology
Premarket evidence — what FDA accepted
source quote (p.4)
“The ECHELON Synergy System is an imaging device and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The MR system produces transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head, body, or extremities. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination.”
source quote (p.10)
“The validation results of the new features using machine learning (DLR, AutoClip, AutoPose Spine, AutoPose Shoulder, and AutoPose Knee) were described below.”
Validation studies (4)
Reader study (MRMC)
n=81 cases
endpoints: signal to noise ratio; sharpness; lesion conspicuity; overall image quality
Retrospective clinical
n=110 cases
endpoints: clinical usability; image quality
Retrospective clinical
n=40 cases
endpoints: performance of AutoClip
Retrospective clinical
n=146 cases
endpoints: time and number of steps in slice positioning
Reported performance (0 observations)
FDA source did not state a quantitative performance metric — non-reporting is itself the signal.
Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.
Predicate network
Postmarket — what happened after clearance
- re_clearance
The FDA AI/ML device list shows a newer 510(k) K241429 (decision 2024-08-13) from FUJIFILM Healthcare Corporation for a matching device line ("ECHELON Synergy MRI System") — a new clearance for the same line is a change event.
first seen 2026-07-08 · k_number:K241429
- recall_reason_pattern
Software/algorithm-related recall in product code LNH (Philips North America, initiated 2026-04-14): "The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction " Recalling firm is another firm in the same product code.
first seen 2026-07-08 · recall res_event_number:98779
- recall_reason_pattern
Software/algorithm-related recall in product code LNH (Philips North America, initiated 2025-12-03): "The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS)." Recalling firm is another firm in the same product code.
first seen 2026-07-08 · recall res_event_number:98111
Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device). Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.
Reimbursement — how devices like this got paid
Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).