BrainInsight

K223268

Hyperfine, Inc. · cleared 2022-12-16 · product code QIH · Radiology

Premarket evidence — what FDA accepted

Device typesamd
source quote (p.4)
BrainInsight is a fully automated MR imaging post-processing medical software that provides image alignment, whole brain segmentation, ventricle segmentation, and midline shift measurements of brain structures from a set MR images.
Algorithmproprietary automated internal pipeline based on machine learning tools
source quote (p.4)
The BrainInsight processing architecture includes a proprietary automated internal pipeline based on machine learning tools.
Adaptive (vs locked)FDA source did not state this
PCCPFDA source did not state this
Cybersecurity addressedFDA source did not state this

Validation studies (1)

Retrospective clinical

sample size not stated

endpoints: Midline Shift Discrepancy; Lateral Ventricle Left Discrepancy; Lateral Ventricle Right Discrepancy; Dice Overlap; Volume Differences

standards: IEC 62304:2006

Reported performance (10 observations)

diceas written: “Dice Overlap (Device) - T1 Left Ventricle85
source quote (p.8)
Application Dice Overlap [%] T1 Device Left Ventricle 85
diceas written: “Dice Overlap (Device) - T1 Right Ventricle83
source quote (p.8)
Application Dice Overlap [%] T1 Device Right Ventricle 83
diceas written: “Dice Overlap (Device) - T1 Whole Brain95
source quote (p.8)
Application Dice Overlap [%] T1 Device Whole Brain 95
diceas written: “Dice Overlap (Device) - T2 Left Ventricle84
source quote (p.8)
Application Dice Overlap [%] T2 Device Left Ventricle 84
diceas written: “Dice Overlap (Device) - T2 Right Ventricle82
source quote (p.8)
Application Dice Overlap [%] T2 Device Right Ventricle 82
diceas written: “Dice Overlap (Device) - T2 Whole Brain96
source quote (p.8)
Application Dice Overlap [%] T2 Device Whole Brain 96
diceas written: “Dice Overlap (Device) - T2-Fast Left Ventricle86
source quote (p.8)
Application Dice Overlap [%] T2-Fast Device Left Ventricle 86
diceas written: “Dice Overlap (Device) - T2-Fast Right Ventricle86
source quote (p.8)
Application Dice Overlap [%] T2-Fast Device Right Ventricle 86
diceas written: “Dice Overlap (Device) - FLAIR Left Ventricle89
source quote (p.8)
Application Dice Overlap [%] FLAIR Device Left Ventricle 89
diceas written: “Dice Overlap (Device) - FLAIR Right Ventricle88
source quote (p.8)
Application Dice Overlap [%] FLAIR Device Right Ventricle 88

Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.

Predicate network

Postmarket — what happened after clearance

0
recalls in product code, 24mo
3
MAUDE reports in code, 12mo
vs code's own 3-yr baseline
0
drift signals on this device

Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device). Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.

Reimbursement — how devices like this got paid

Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).

RIGOR™ Precedent · public FDA/CMS data · descriptive decision-support, not regulatory or reimbursement advice. Share this page: radar.healthai.com/precedent/device/K223268