Annalise Enterprise CTB Triage Trauma

K223240

Annalise-AI Pty Ltd · cleared 2023-04-03 · product code QAS · Radiology

Premarket evidence — what FDA accepted

Device typesamd
source quote (p.6)
Annalise Enterprise CTB Triage Trauma is a software workflow tool which uses an artificial intelligence (AI) algorithm to identify suspected findings on non-contrast brain CT studies in the medical care environment.
AlgorithmAI algorithm a convolutional neural network trained using deep-learning techniques
source quote (p.6)
Radiological findings are identified by the device using an AI algorithm a convolutional neural network trained using deep-learning techniques.
Adaptive (vs locked)FDA source did not state this
PCCPFDA source did not state this
Cybersecurity addressedFDA source did not state this

Validation studies (1)

Retrospective clinical

n=3,363 cases · 5 site(s)

endpoints: Sensitivity; Specificity; Triage effectiveness (turn-around time)

Reported performance (1 observation)

time_to_resultas written: “Triage turn-around time81.6CI 95% CI: 80.3 – 82.9
source quote (p.10)
The results demonstrated a triage turn-around time of 81.6 (95% CI: 80.3 – 82.9) seconds

Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.

Predicate network

Postmarket — what happened after clearance

0
recalls in product code, 24mo
0
MAUDE reports in code, 12mo
vs code's own 3-yr baseline
3
drift signals on this device
  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K231767 (decision 2023-09-22) from Annalise-AI Pty Ltd for a matching device line ("Annalise Enterprise CTB Triage Trauma") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K231767

  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K231384 (decision 2023-09-22) from Annalise-AI Pty Ltd. for a matching device line ("Annalise Enterprise CTB Triage Trauma") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K231384

  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K231094 (decision 2023-08-15) from Annalise-AI Pty Ltd for a matching device line ("Annalise Enterprise CTB Triage-OH") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K231094

Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device). Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.

Reimbursement — how devices like this got paid

Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).

RIGOR™ Precedent · public FDA/CMS data · descriptive decision-support, not regulatory or reimbursement advice. Share this page: radar.healthai.com/precedent/device/K223240