BoneView
K222176Gleamer · cleared 2023-03-02 · product code QBS · Radiology
Premarket evidence — what FDA accepted
source quote (p.5)
“BoneView 1.1-US is a software-only device intended to assist clinicians in the interpretation of: limbs radiographs of children/adolescents and limbs, pelvis, rib cage, and dorsolumbar vertebra radiographs of adults.”
source quote (p.9)
“Machine Learning Methodology Supervised Deep Learning”
source quote (p.6)
“These two operating points are implemented in the User Interface as follow: Dotted-line Bounding Box: suspicious area / subtle fracture (when the level of confidence of the AI algorithm associated with the finding is above “high-sensitivity operating point" and below “high-specificity operating point") displayed as a dotted bounding box around the area of interest Solid-line Bounding Box: definite or unequivocal fractures (when the level of confidence of the AI algorithm associated with the finding is above “high-specificity operating point") displayed as a solid bounding box around the area of interest”
Validation studies (3)
Standalone
n=2,000 images
endpoints: Sensitivity; Specificity
Standalone
n=8,918 images
endpoints: Sensitivity; Specificity
Retrospective clinical
n=480 cases
endpoints: diagnostic accuracy of readers aided by BoneView is superior to the diagnostic accuracy of readers unaided by BoneView as determined by the Specificity/Sensitivity pair (primary endpoint)
Reported performance (8 observations)
source quote (p.12)
“Sensitivity (with 95% Clopper-Pearson CI) and Specificity (with 95% Clopper-Pearson CI) of BoneView at the examination-level at the high-sensitivity operating point and high-specificity operating point on the children/adolescents clinical performance study dataset ... Global n(positive)= 1,000 n(negative)= 1,000 ... Sensitivity - 95% Clopper-Pearson CI 0.909 [0.889 - 0.926]”
source quote (p.12)
“Specificity (with 95% Clopper-Pearson CI) and Sensitivity (with 95% Clopper-Pearson CI) of BoneView at the examination-level at the high-sensitivity operating point and high-specificity operating point on the children/adolescents clinical performance study dataset ... Global n(positive)= 1,000 n(negative)= 1,000 ... Specificity - 95% Clopper-Pearson CI 0.821 [0.796 - 0.844]”
source quote (p.12)
“Sensitivity - 95% Clopper-Pearson Cl 0.792 [0.766 - 0.817]”
source quote (p.12)
“Specificity - 95% Clopper-Pearson CI 0.965 [0.952 - 0.976]”
source quote (p.14)
“Sensitivity (with 95% Clopper-Pearson CI) and Specificity (with 95% Clopper-Pearson CI) of BoneView at the examination-level at the high-sensitivity operating point and high-specificity operating point on the merged datasets ... Global n(positive)= 3,886 n(negative)= 5,032 ... Sensitivity - 95% Clopper-Pearson CI 0.928 [0.919 - 0.936]”
source quote (p.14)
“Specificity (with 95% Clopper-Pearson CI) and Sensitivity (with 95% Clopper-Pearson CI) of BoneView at the examination-level at the high-sensitivity operating point and high-specificity operating point on the merged datasets ... Global n(positive)= 3,886 n(negative)= 5,032 ... Specificity - 95% Clopper-Pearson CI 0.811 [0.8 - 0.821]”
source quote (p.14)
“Sensitivity (with 95% Clopper-Pearson CI) and Specificity (with 95% Clopper-Pearson CI) of BoneView at the examination-level at the high-sensitivity operating point and high-specificity operating point on the merged datasets ... Global n(positive)= 3,886 n(negative)= 5,032 ... Sensitivity - 95% Clopper-Pearson CI 0.841 [0.829 - 0.853]”
source quote (p.14)
“Specificity (with 95% Clopper-Pearson CI) and Sensitivity (with 95% Clopper-Pearson CI) of BoneView at the examination-level at the high-sensitivity operating point and high-specificity operating point on the merged datasets ... Global n(positive)= 3,886 n(negative)= 5,032 ... Specificity - 95% Clopper-Pearson CI 0.932 [0.925 - 0.939]”
Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.
Predicate network
Postmarket — what happened after clearance
Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device). Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.
Reimbursement — how devices like this got paid
Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).