DTX Studio Clinic 3.0
K221921Nobel Biocare AB · cleared 2023-03-28 · product code MYN · Radiology
Premarket evidence — what FDA accepted
source quote (p.3)
“DTX Studio Clinic is a computer assisted detection (CADe) device that analyses intraoral radiographs to identify and localize dental findings, which include caries, calculus, periapical radiolucency, root canal filling deficiency, discrepancy at margin of an existing restoration and bone loss.”
source quote (p.5)
“DTX Studio Clinic features an Al-powered Focus Area Detection algorithm which analyzes intraoral radiographs for potential dental findings or image artifacts.”
Validation studies (2)
Standalone
n=452 images
endpoints: overall sensitivity; sensitivity per dental finding type; overall precision; localization accuracy (IoU); Dice score
standards: 21 CFR § 820, ISO 13485:2016, IEC 62304:2006+A1:2015, ISO 14971:2019
Reader study (MRMC)
n=216 images
endpoints: Area Under the Curve (AUC) increase; AUC difference per dental finding; sensitivity on instance level; specificity at image level
standards: 21 CFR § 820, ISO 13485:2016, IEC 62304:2006+A1:2015, ISO 14971:2019
Reported performance (34 observations)
source quote (p.10)
“Overall precision was found to be 0.45 (with 95% CI [0.40, 0.50]).”
source quote (p.10)
“0.70 [0.65, 0.75]”
source quote (p.10)
“58.6 [56.2, 60.9]”
source quote (p.10)
“71.9 [69.9, 74.0]”
source quote (p.10)
“0.68 [0.59, 0.77]”
source quote (p.10)
“48.9 [44.9, 52.9]”
source quote (p.10)
“63.7 [59.9, 67.5]”
source quote (p.11)
“0.95 [0.91, 0.99]”
source quote (p.11)
“51.9 [49.3, 54.6]”
source quote (p.11)
“66.9 [64.3, 69.4]”
source quote (p.11)
“0.82 [0.77, 0.87]”
source quote (p.11)
“48.4 [46.0, 50.7]”
source quote (p.11)
“63.5 [61.3, 65.8]”
source quote (p.11)
“0.78 [0.75, 0.81]”
source quote (p.11)
“44.8 [43.4, 46.3]”
source quote (p.11)
“60.1 [58.7, 61.6]”
source quote (p.11)
“0.80 [0.76, 0.84]”
source quote (p.11)
“55.5 [53.7, 57.3]”
source quote (p.11)
“70.1 [68.4, 71.7]”
source quote (p.11)
“overall increase for all dental finding types together of 8.7% (CI [6.5, 10.9]).”
source quote (p.11)
“an increase of 13.5% for root canal filling deficiency”
source quote (p.11)
“10.2% for periapical radiolucency”
source quote (p.11)
“10.1% for discrepancy at margin of an existing restoration”
source quote (p.11)
“7.2% for calculus”
source quote (p.11)
“6.1% for caries”
source quote (p.11)
“5.6% for bone loss”
source quote (p.12)
“overall by 22.4% (with CI [20.1, 24.7])”
source quote (p.12)
“19.6% for caries”
source quote (p.12)
“23.5% for bone loss”
source quote (p.12)
“18.1% for calculus”
source quote (p.12)
“28.5% for discrepancy at restoration margin”
source quote (p.12)
“20.6% for periapical radiolucency”
source quote (p.12)
“27.4% for root canal filling deficiency”
source quote (p.12)
“overall by 8.7 % with CI [6.6, 10.7]”
Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.
Predicate network
Postmarket — what happened after clearance
- re_clearance
The FDA AI/ML device list shows a newer 510(k) K231898 (decision 2023-12-08) from Nobel Biocare AB for a matching device line ("DTX Studio Clinic (4.0)") — a new clearance for the same line is a change event.
first seen 2026-07-08 · k_number:K231898
Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device). Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.
Reimbursement — how devices like this got paid
Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).