BriefCase
K221314Aidoc Medical, Ltd. · cleared 2022-06-03 · product code QAS · Radiology
Premarket evidence — what FDA accepted
source quote (p.3)
“BriefCase is a radiological computer aided triage and notification software indicated for use in the analysis of head CTA images in adults or transitional adolescents aged 18 and older.”
source quote (p.5)
“Both devices are artificial intelligence, deep-learning algorithms incorporated software packages for use with DICOM 3.0 compliant CT scanners, PACS, and radiology workstations.”
source quote (p.5)
“Both devices are intended to aid in triage and prioritization of radiological images and utilize the same design of deep learning algorithm trained on medical images.”
Validation studies (1)
Retrospective clinical
n=383 patients
endpoints: evaluate the software's performance in identifying head CTA images containing M1 Large Vessel Occlusion (M1-LVO)
Reported performance (5 observations)
source quote (p.7)
“Sensitivity was 88.8% (95% CI: 81.9%, 93.8%)”
source quote (p.7)
“and specificity was 87.2% (95% CI: 82.5%, 91.1%).”
source quote (p.8)
“NPV was 99.5% (95% CI: 99.3%-99.8%)”
source quote (p.8)
“and PPV was 20.1% (95% CI: 13.2%-24.2%).”
source quote (p.8)
“The BriefCase mean time-to-notification for M1-LVO was 3.8 minutes (95% CI: 3.6-4.0).”
Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.
Predicate network
Postmarket — what happened after clearance
- re_clearance
The FDA AI/ML device list shows a newer 510(k) K253578 (decision 2026-02-26) from Aidoc Medical , Ltd. for a matching device line ("BriefCase-Triage: CARE Multi-Triage CT for Pneumothorax; Pericardial effusion; Large aortic aneurysm; Shoulder fracture or dislocation device") — a new clearance for the same line is a change event.
first seen 2026-07-08 · k_number:K253578
- re_clearance
The FDA AI/ML device list shows a newer 510(k) K251195 (decision 2026-01-27) from Aidoc Medical , Ltd. for a matching device line ("BriefCase-Triage") — a new clearance for the same line is a change event.
first seen 2026-07-08 · k_number:K251195
- re_clearance
The FDA AI/ML device list shows a newer 510(k) K252970 (decision 2026-01-07) from Aidoc Medical , Ltd. for a matching device line ("BriefCase-Triage: CARE Multi-triage CT Body") — a new clearance for the same line is a change event.
first seen 2026-07-08 · k_number:K252970
- re_clearance
The FDA AI/ML device list shows a newer 510(k) K253265 (decision 2025-11-06) from Aidoc Medical , Ltd. for a matching device line ("BriefCase-Triage") — a new clearance for the same line is a change event.
first seen 2026-07-08 · k_number:K253265
- re_clearance
The FDA AI/ML device list shows a newer 510(k) K251406 (decision 2025-05-30) from Aidoc Medical, Ltd. for a matching device line ("BriefCase-Triage") — a new clearance for the same line is a change event.
first seen 2026-07-08 · k_number:K251406
- re_clearance
The FDA AI/ML device list shows a newer 510(k) K250248 (decision 2025-02-14) from Aidoc Medical, Ltd. for a matching device line ("BriefCase-Triage") — a new clearance for the same line is a change event.
first seen 2026-07-08 · k_number:K250248
- …and 15 more.
Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device). Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.
Reimbursement — how devices like this got paid
Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).