Venue Fit
K220848GE Medical Systems · cleared 2022-06-27 · product code IYN · Radiology
Premarket evidence — what FDA accepted
source quote (p.8)
“cNerve: may help the user to detect and track nerves during the scouting stage of a nerve block procedure, prior to inserting the needle to inject the anesthetic material. Is similar to the Nerve Track tool on cleared reference HS40 system by Samsung (K210426). The cNerve tool on proposed Venue Fit system is available for adults only. Expanding B-lines tool for pediatrics (excluding neonates). This allows Lung sweep and Lung Diagram to pediatrics (excluding neonates) as well. B-lines tool is available on the cleared predicate Venue Fit (K203137) for adults only.”
source quote (p.8)
“cNerve: may help the user to detect and track nerves during the scouting stage of a nerve block procedure, prior to inserting the needle to inject the anesthetic material. Is similar to the Nerve Track tool on cleared reference HS40 system by Samsung (K210426). The cNerve tool on proposed Venue Fit system is available for adults only. Expanding B-lines tool for pediatrics (excluding neonates). This allows Lung sweep and Lung Diagram to pediatrics (excluding neonates) as well.”
Validation studies (1)
Retrospective clinical
n=44 patients
endpoints: Sequence accuracy requirement – for testing overall cNerve performance: At least 70% of the sequences are meaningfully detected; At least 80% of the meaningfully detected sequences are successfully detected (meeting frame accuracy criteria)
Reported performance (2 observations)
source quote (p.10)
“At least 70% of the sequences are meaningfully detected”
source quote (p.10)
“and at least 80% of the meaningfully detected sequences are successfully detected (meeting frame accuracy criteria).”
Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.
Predicate network
Postmarket — what happened after clearance
- recall_reason_pattern
Software/algorithm-related recall in product code IYN (GE Medical Systems, LLC, initiated 2025-09-18): "The Ultrasound-Guided Attenuation Parameter (UGAP) measurement data may display inaccurate values representing liver steatosis. This could potentially lead to inappropriate clinica" Recalling firm matches this device's applicant.
first seen 2026-07-08 · recall res_event_number:97726
- recall_reason_pattern
Software/algorithm-related recall in product code IYN (GE Medical Systems China Co., Ltd. Dev. Zone National Hi-Tech; No., initiated 2025-05-16): "GE HealthCare has become aware that the Estimated Fetal Weight (EFW) measurement data feature on the Versana Premier R3 and LOGIQ F R3 series ultrasound systems can display previou" Recalling firm matches this device's applicant.
first seen 2026-07-08 · recall res_event_number:96992
- re_clearance
The FDA AI/ML device list shows a newer 510(k) K251322 (decision 2025-07-25) from GE Medical Systems Ultrasound and Primary care Diagnostics, for a matching device line ("Venue; Venue Go; Venue Fit; Venue Sprint") — a new clearance for the same line is a change event.
first seen 2026-07-08 · k_number:K251322
- recall_reason_pattern
Software/algorithm-related recall in product code IYN (Philips Ultrasound, LLC, initiated 2025-10-31): "Ultrasound system compatibility issues with Apple devices running iOS 18 may cause a failure to perform live imagining." Recalling firm is another firm in the same product code.
first seen 2026-07-08 · recall res_event_number:97843
- recall_reason_pattern
Software/algorithm-related recall in product code IYN (Siemens Medical Solutions USA, Inc., initiated 2024-08-15): "If ultrasound systems with software, are changed from factory default to : 1) Milliliters per second (ml/sec, mL/sec) or 2) Milliliters per minute (ml/min, mL/min); then systems wi" Recalling firm is another firm in the same product code.
first seen 2026-07-08 · recall res_event_number:95254
Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device). Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.
Reimbursement — how devices like this got paid
Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).