BriefCase
K220709Aidoc Medical, Ltd. · cleared 2022-10-07 · product code QAS · Radiology
Premarket evidence — what FDA accepted
source quote (p.3)
“BriefCase is a radiological computer aided triage and notification software indicated for use in the analysis of head CTA images in adults or transitional adolescents aged 18 and older. The device is intended to assist hospital networks and appropriately trained medical specialists in workflow triage by flagging and communication of suspected positive findings of complete Large Vessel Occlusion (LVO) - MCA-M1, PCA-P1, ACA-A1, ICA, Basilar; and Medium Vessel Occlusions (MeVO) - MCA-M2, MCA-proximal M3, PCA-P2, PCA-proximal P3, ACA-A2, ACA-proximal A3, and Vertebral-V4. BriefCase uses an artificial intelligence algorithm to analyze images and highlight cases with detected findings on a standalone desktop application in parallel to the ongoing standard of care image interpretation.”
source quote (p.5)
“Both devices are artificial intelligence, deep-learning algorithms incorporated software packages for use with DICOM 3.0 compliant CT scanners, PACS, and radiology workstations.”
source quote (p.5)
“Both devices are intended to aid in triage and prioritization of radiological images and utilize the same design of deep learning algorithm trained on medical images.”
Validation studies (1)
Retrospective clinical
n=342 cases · 5 site(s)
endpoints: sensitivity; specificity; BriefCase time-to-notification compared to the predicate device; Positive Predictive Value (PPV); Negative Predictive Value (NPV); Positive Likelihood Ratio (PLR); Negative Likelihood Ratio (NLR)
Reported performance (8 observations)
source quote (p.8)
“Sensitivity was 91.3% (95% CI: 83.6%, 96.2%)”
source quote (p.8)
“specificity was 85.6% (95% CI: 80.6%, 89.7%).”
source quote (p.8)
“The BriefCase mean time-to-notification for VO was 2.23 minutes (95% CI: 2.22-2.23).”
source quote (p.8)
“The time-to-notification for the predicate BriefCase LVO was 3.80 minutes (95% CI: 3.60-4.00).”
source quote (p.9)
“NPV was 98.9% (95% CI: 4.66%- 99.4%)”
source quote (p.9)
“PPV was 41.3% (95% CI: 34.1%- 49.0%).”
source quote (p.9)
“PLR was 6.34% (95% CI: 97.9%- 8.63%)”
source quote (p.9)
“NLR was 0.10% (95% CI: 0.05%- 0.20%).”
Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.
Predicate network
Postmarket — what happened after clearance
- re_clearance
The FDA AI/ML device list shows a newer 510(k) K253578 (decision 2026-02-26) from Aidoc Medical , Ltd. for a matching device line ("BriefCase-Triage: CARE Multi-Triage CT for Pneumothorax; Pericardial effusion; Large aortic aneurysm; Shoulder fracture or dislocation device") — a new clearance for the same line is a change event.
first seen 2026-07-08 · k_number:K253578
- re_clearance
The FDA AI/ML device list shows a newer 510(k) K251195 (decision 2026-01-27) from Aidoc Medical , Ltd. for a matching device line ("BriefCase-Triage") — a new clearance for the same line is a change event.
first seen 2026-07-08 · k_number:K251195
- re_clearance
The FDA AI/ML device list shows a newer 510(k) K252970 (decision 2026-01-07) from Aidoc Medical , Ltd. for a matching device line ("BriefCase-Triage: CARE Multi-triage CT Body") — a new clearance for the same line is a change event.
first seen 2026-07-08 · k_number:K252970
- re_clearance
The FDA AI/ML device list shows a newer 510(k) K253265 (decision 2025-11-06) from Aidoc Medical , Ltd. for a matching device line ("BriefCase-Triage") — a new clearance for the same line is a change event.
first seen 2026-07-08 · k_number:K253265
- re_clearance
The FDA AI/ML device list shows a newer 510(k) K251406 (decision 2025-05-30) from Aidoc Medical, Ltd. for a matching device line ("BriefCase-Triage") — a new clearance for the same line is a change event.
first seen 2026-07-08 · k_number:K251406
- re_clearance
The FDA AI/ML device list shows a newer 510(k) K250248 (decision 2025-02-14) from Aidoc Medical, Ltd. for a matching device line ("BriefCase-Triage") — a new clearance for the same line is a change event.
first seen 2026-07-08 · k_number:K250248
- …and 12 more.
Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device). Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.
Reimbursement — how devices like this got paid
Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).