Saige-Dx
K220105DeepHealth, Inc. · cleared 2022-05-12 · product code QDQ · Radiology
Premarket evidence — what FDA accepted
source quote (p.5)
“Saige-Dx is a software device that processes screening mammograms using artificial intelligence to aid interpreting radiologists.”
source quote (p.5)
“Saige-Dx is a software device that processes screening mammograms using artificial intelligence to aid interpreting radiologists.”
source quote (p.6)
“DeepHealth ensured that there was no overlap between the data used to train and test the Saige-Dx AI algorithm.”
Validation studies (2)
Reader study (MRMC)
n=240 cases
endpoints: compare breast cancer detection performance of radiologists reading with the aid of Saige-Dx versus without Saige-Dx
Retrospective clinical
n=1,304 cases · 9 site(s)
Reported performance (2 observations)
source quote (p.7)
“Saige-Dx exhibited an AUC of 0.930 (95% CI: 0.902, 0.958) on the dataset”
source quote (p.7)
“The reader performance increased with the aid of Saige-Dx from an average AUC of 0.865 when unaided to 0.925 when aided (difference of 0.06; 95% CI: 0.041, 0.079, p < 0.00001).”
Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.
Predicate network
Postmarket — what happened after clearance
- re_clearance
The FDA AI/ML device list shows a newer 510(k) K251873 (decision 2025-08-11) from DeepHealth, Inc. for a matching device line ("Saige-Dx") — a new clearance for the same line is a change event.
first seen 2026-07-08 · k_number:K251873
- re_clearance
The FDA AI/ML device list shows a newer 510(k) K243705 (decision 2024-12-19) from DeepHealth, Inc for a matching device line ("Saige-Density (2.5.0)") — a new clearance for the same line is a change event.
first seen 2026-07-08 · k_number:K243705
- re_clearance
The FDA AI/ML device list shows a newer 510(k) K243688 (decision 2024-12-19) from DeepHealth, Inc. for a matching device line ("Saige-Dx (3.1.0)") — a new clearance for the same line is a change event.
first seen 2026-07-08 · k_number:K243688
- re_clearance
The FDA AI/ML device list shows a newer 510(k) K241747 (decision 2024-11-18) from DeepHealth, Inc for a matching device line ("Saige-Dx") — a new clearance for the same line is a change event.
first seen 2026-07-08 · k_number:K241747
- re_clearance
The FDA AI/ML device list shows a newer 510(k) K222275 (decision 2022-12-16) from DeepHealth, Inc. for a matching device line ("Saige-Density") — a new clearance for the same line is a change event.
first seen 2026-07-08 · k_number:K222275
Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device). Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.
Reimbursement — how devices like this got paid
Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).