CardIQ Suite

K213725

GE Medical Systems SCS · cleared 2022-03-11 · product code JAK · Radiology

Premarket evidence — what FDA accepted

Device typesamd
source quote (p.3)
CardIQ Suite is a non-invasive software application designed to provide an optimized application to analyze cardiovascular anatomy and pathology based on 2D or 3D CT cardiac non contrast and angiography DICOM data from acquisitions of the heart.
Algorithmdeep learning algorithm
source quote (p.6)
The calcium segmentation/labeling is using a new deep learning algorithm.
Adaptive (vs locked)FDA source did not state this
PCCPNo
Cybersecurity addressedNo

Validation studies (2)

Bench

sample size not stated

endpoints: algorithm's capability to automatically segment, label and score the calcific regions in the coronary arteries

standards: 21CFR 820, ISO 13485, NEMA PS 3.1 - 3.20 (2016) Digital Imaging and Communications in Medicine (DICOM) Set (Radiology) standard

Reader study (MRMC)

sample size not stated

endpoints: correlation between manual and automated methods for computing the total calcium score; equivalent performance of the CardIQ Suite software to the predicate device SmartScore 4.0

Reported performance (0 observations)

FDA source did not state a quantitative performance metric — non-reporting is itself the signal.

Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.

Predicate network

Postmarket — what happened after clearance

75
recalls in product code, 24mo
192
MAUDE reports in code, 12mo
-22%
vs code's own 3-yr baseline
11
drift signals on this device
  • recall_reason_pattern

    Software/algorithm-related recall in product code JAK (GE Medical Systems, LLC, initiated 2026-03-26): "GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with " Recalling firm matches this device's applicant.

    first seen 2026-07-08 · recall res_event_number:98738

  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K243672 (decision 2025-06-18) from GE Medical Systems SCS for a matching device line ("CardIQ Suite") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K243672

  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K233731 (decision 2024-08-01) from GE Medical Systems SCS for a matching device line ("CardIQ Suite") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K233731

  • recall_reason_pattern

    Software/algorithm-related recall in product code JAK (PHILIPS MEDICAL SYSTEMS, initiated 2026-03-07): "Philips has identified three software issues: 1. During a continuous CT (CCT) scan, there is the potential that the Gantry could remain at the current scan position after pressing" Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:98588

  • recall_reason_pattern

    Software/algorithm-related recall in product code JAK (Siemens Medical Solutions USA, Inc, initiated 2025-12-19): "To remove the software applications from certain CT systems as the applications have not received FDA 510(k) clearance." Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:98206

  • recall_reason_pattern

    Software/algorithm-related recall in product code JAK (PHILIPS MEDICAL SYSTEMS, initiated 2025-09-25): "Issue 1: The potential for unintentional continued gantry/couch movement when a specific button series is used requiring use of manual stop. Issue 2. When performing a helical/Axia" Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:97699

  • …and 5 more.