AI-Rad Companion Brain MR

K213706

Siemens Healthcare GmBh · cleared 2022-04-15 · product code QIH · Radiology

Premarket evidence — what FDA accepted

Device typesamd
source quote (p.3)
AI-Rad Companion Brain MR is a post-processing image analysis software that assists clinicians in viewing, analyzing, and evaluating MR brain images.
Algorithmwhite matter hyperintensities segmentation and analysis algorithm
source quote (p.10)
To validate the AI-Rad Companion Brain MR software from clinical perspective, the white matter hyperintensities segmentation and analysis algorithm underwent a scientific evaluation.
Adaptive (vs locked)FDA source did not state this
PCCPFDA source did not state this
Cybersecurity addressedYes
source quote (p.10)
Siemens Healthineers adheres to the cybersecurity requirements as defined the FDA Guidance “Content of Premarket Submissions for Management for Cybersecurity in Medical Devices,” issued October 2, 2014 by implementing a process of preventing unauthorized access, modifications, misuse or denial of use, or the unauthorized use of information that is stored, accessed, or transferred from a medical device to an external recipient.

Validation studies (3)

Bench

sample size not stated

endpoints: test the functionality of AI-Rad Companion Brain MR; assess the performance claims; assess the claim of substantial equivalence to the predicate device

Retrospective clinical

n=64 patients · 7 site(s)

endpoints: PCC 95% Confidence Interval includes 0.91; ICC 95% Confidence Interval includes 0.95; Mean Dice score >= 0.58; Mean F1-score >= 0.57

Retrospective clinical

n=25 patients · 7 site(s)

endpoints: Lower Bound of the 95% Bootstrap CI Dice >= 0.63

Reported performance (4 observations)

agreement_kappaas written: “ICC (Volumetric Segmentation)0.97CI [0.96,0.98]
source quote (p.11)
AVG 0.97 95% CI [0.96,0.98]
diceas written: “Dice (Voxel-wise Segmentation)0.6CI [0.53,0.63]
source quote (p.11)
AVG 0.60 95% CI [0.53,0.63]
f1as written: “F1-score (WMH Lesion-wise Segmentation)0.6CI [0.57,0.64]
source quote (p.11)
AVG 0.60 95% CI [0.57,0.64]
diceas written: “Dice (Reproducibility)0.79CI [0.77,0.81]
source quote (p.11)
AVG 0.79 95% CI [0.77,0.81]

Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.

Predicate network

Postmarket — what happened after clearance

0
recalls in product code, 24mo
3
MAUDE reports in code, 12mo
vs code's own 3-yr baseline
1
drift signals on this device
  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K253057 (decision 2026-01-22) from Siemens Healthcare GmbH for a matching device line ("AI-Rad Companion Brain MR") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K253057

Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device). Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.

Reimbursement — how devices like this got paid

Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).