SIGNA Hero
K213668GE Healthcare (GE Medical Systems, LLC) · cleared 2022-01-20 · product code LNH · Radiology
Premarket evidence — what FDA accepted
source quote (p.3)
“The SIGNA Hero is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body.”
Validation studies (2)
Bench
sample size not stated
standards: ANSI/AAMI ES60601-1, IEC 60601-1-2, IEC 60601-2-33, IEC 62304, ISO 10993-1, IEC 62464-1, NEMA MS 3, NEMA MS 4, NEMA MS 8, NEMA PS3 standard for DICOM
Reader study (MRMC)
sample size not stated
Reported performance (0 observations)
FDA source did not state a quantitative performance metric — non-reporting is itself the signal.
Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.
Predicate network
Postmarket — what happened after clearance
- recall_reason_pattern
Software/algorithm-related recall in product code LNH (Philips North America, initiated 2026-04-14): "The potential for stiffness value errors when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm, leading to the reconstruction " Recalling firm is another firm in the same product code.
first seen 2026-07-08 · recall res_event_number:98779
- recall_reason_pattern
Software/algorithm-related recall in product code LNH (Philips North America, initiated 2025-12-03): "The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS)." Recalling firm is another firm in the same product code.
first seen 2026-07-08 · recall res_event_number:98111
Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device). Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.
Reimbursement — how devices like this got paid
Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).