Aorta-CAD
K213353Imagen Technologies, Inc · cleared 2022-09-20 · product code MYN · Radiology
Premarket evidence — what FDA accepted
source quote (p.5)
“Aorta-CAD is computer-assisted detection (CADe) software device that analyzes chest radiograph studies for suspicious regions of interest (ROIs).”
source quote (p.5)
“The device uses a deep learning algorithm to identify ROIs and produce boxes around the ROIs. Aorta-CAD uses modern deep learning and computer vision techniques to analyze chest radiographs.”
source quote (p.7)
“HIPAA Compliant”
Validation studies (2)
Standalone
n=5,000 cases
endpoints: sensitivity; specificity; Area Under the Curve (AUC) of the Receiver Operating Characteristic (ROC) curve
Reader study (MRMC)
n=244 cases
endpoints: accuracy of readers aided by Aorta-CAD superior to unaided readers per category as determined by the case-level Area Under the Curve (AUC) of the Receiver Operating Characteristic (ROC) curve
Reported performance (7 observations)
source quote (p.8)
“The results of the standalone testing demonstrated that Aorta-CAD detects ROIs with high sensitivity (0.910; 95% Wilson's Confidence Interval: 0.896, 0.922)”
source quote (p.8)
“high specificity (0.896; 95% Wilson's Confidence Interval: 0.889, 0.902)”
source quote (p.8)
“high Area Under the Curve (AUC) of the Receiver Operating Characteristic (ROC) curve (0.974, 95% Bootstrap Confidence Interval: 0.971, 0.977).”
source quote (p.10)
“Positive Predictive Value 0.812 (0.794, 0.829)”
source quote (p.10)
“Negative Predictive Value 0.958 (0.951, 0.965)”
source quote (p.10)
“Positive Predictive Value 0.296 (0.263, 0.331)”
source quote (p.10)
“Negative Predictive Value 0.990 (0.987, 0.993)”
Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.
Predicate network
Postmarket — what happened after clearance
Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device). Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.
Reimbursement — how devices like this got paid
Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).