Autoplaque

K212758

Cedars-Sinai Medical Center: AIM · cleared 2023-05-19 · product code LLZ · Radiology

Premarket evidence — what FDA accepted

Device typesamd
source quote (p.3)
Autoplaque is a workstation-based post processing application. It is a non-invasive diagnostic reading software intended for use by cardiologists and radiologists as an interactive tool for viewing and analyzing cardiac CT data for determining the presence and extent of coronary plaques and luminal stenoses.
AlgorithmDeep learning-based vessel, plaque and lumen segmentation
source quote (p.6)
Deep learning-based vessel, plaque and lumen segmentation, which is reviewed and can be edited, if necessary, by the clinician.
Adaptive (vs locked)No
source quote (p.6)
Deep learning-based vessel, plaque and lumen segmentation, which is reviewed and can be edited, if necessary, by the clinician.
PCCPNo
Cybersecurity addressedNo

Validation studies (3)

Retrospective clinical

n=201 patients

endpoints: total plaque volume; calcified plaque volume; non-calcified plaque volume; diameter stenosis; intraclass correlation coefficient; correlation coefficient

Retrospective clinical

n=175 patients

endpoints: total plaque volume; calcified plaque volume; non-calcified plaque volume; diameter stenosis; intraclass correlation coefficient; correlation coefficient

Standalone

n=27 patients

endpoints: total plaque volume; non-calcified volume; calcified plaque volume; diameter stenosis; intraclass correlation agreement

Reported performance (0 observations)

FDA source did not state a quantitative performance metric — non-reporting is itself the signal.

Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.

Predicate network

Postmarket — what happened after clearance

48
recalls in product code, 24mo
295
MAUDE reports in code, 12mo
+683%
vs code's own 3-yr baseline
26
drift signals on this device
  • adverse_event_inflection

    MAUDE adverse-event reports for product code LLZ: 295 in the 12 months ending 2026-06, vs a 37.7/12mo average over the prior 3 windows (+683%). Code-level count — reports are not attributed to this specific device.

    first seen 2026-07-08 · openFDA /device/event.json count=date_received product_code=LLZ

  • recall_reason_pattern

    Software/algorithm-related recall in product code LLZ (GE Medical Systems SCS, initiated 2026-05-08): "GE HealthCare has become aware of a context synchronization issue in AW Server 3.2 ext. 6.5. When a user selects a patient or exam in the AW Server Web Client worklist and launches" Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:99042

  • recall_reason_pattern

    Software/algorithm-related recall in product code LLZ (PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Veenpluis 4-6 Best Netherlands, initiated 2026-03-05): "Potential that mis-ordered frames in Vue Motion during dynamic cine runs may cause images frames to display out of sequence." Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:98520

  • recall_reason_pattern

    Software/algorithm-related recall in product code LLZ (GE Medical Systems, LLC, initiated 2026-01-30): "There is a potential cybersecurity vulnerability affecting certain versions of Centricity Universal Viewer. User login credentials may be exposed on the local client workstation, w" Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:98428

  • recall_reason_pattern

    Software/algorithm-related recall in product code LLZ (Focalyx Technologies, LLC., initiated 2025-12-23): "Software device that is an accessory for image-guided interventional and diagnostic procedures involving prostate has accessories that may not function properly with Windows 10, wh" Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:98137

  • recall_reason_pattern

    Software/algorithm-related recall in product code LLZ (DICOM Grid, Inc., initiated 2025-11-18): "Software intended to aid in diagnosing conditions, planning treatments, visualizing anatomical structures has a bug that, if all of following are met: Viewing images in InteleShare" Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:97993

  • …and 20 more.

Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device). Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.

Reimbursement — how devices like this got paid

Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).

RIGOR™ Precedent · public FDA/CMS data · descriptive decision-support, not regulatory or reimbursement advice. Share this page: radar.healthai.com/precedent/device/K212758