Philips EPIQ Diagnostic Ultrasound System, Philips Affiniti Diagnostic Ultrasound System
K212704Philips Medical Systems · cleared 2021-09-24 · product code IYN · Radiology
Premarket evidence — what FDA accepted
source quote (p.6)
“The purpose of this Traditional 510(k) Pre-Market Notification is to introduce the following two new Liver Fat Quantification software features:”
source quote (p.8)
“The Liver Fat Quantification software feature functionalities allow the end user to perform the following measurements during an ultrasound exam: (1) Attenuation Quantification: a measure of the attenuation coefficient in dB/cm-MHz of a region of interest in the liver (2) Hepato-Renal Index (HRI) Quantification: a ratio measure of the echogenicity (‘brightness') of the liver parenchyma relative to that of the adjacent renal cortex”
Validation studies (1)
Bench
sample size not stated
standards: FDA's Guidance for Content of Premarket Submissions for Software Contained in Medical Devices issued May 11, 2005
Reported performance (0 observations)
FDA source did not state a quantitative performance metric — non-reporting is itself the signal.
Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.
Predicate network
Postmarket — what happened after clearance
- recall_reason_pattern
Software/algorithm-related recall in product code IYN (GE Medical Systems, LLC, initiated 2025-09-18): "The Ultrasound-Guided Attenuation Parameter (UGAP) measurement data may display inaccurate values representing liver steatosis. This could potentially lead to inappropriate clinica" Recalling firm matches this device's applicant.
first seen 2026-07-08 · recall res_event_number:97726
- recall_reason_pattern
Software/algorithm-related recall in product code IYN (Philips Ultrasound, LLC, initiated 2025-10-31): "Ultrasound system compatibility issues with Apple devices running iOS 18 may cause a failure to perform live imagining." Recalling firm is another firm in the same product code.
first seen 2026-07-08 · recall res_event_number:97843
- recall_reason_pattern
Software/algorithm-related recall in product code IYN (GE Medical Systems China Co., Ltd. Dev. Zone National Hi-Tech; No., initiated 2025-05-16): "GE HealthCare has become aware that the Estimated Fetal Weight (EFW) measurement data feature on the Versana Premier R3 and LOGIQ F R3 series ultrasound systems can display previou" Recalling firm is another firm in the same product code.
first seen 2026-07-08 · recall res_event_number:96992
- recall_reason_pattern
Software/algorithm-related recall in product code IYN (Siemens Medical Solutions USA, Inc., initiated 2024-08-15): "If ultrasound systems with software, are changed from factory default to : 1) Milliliters per second (ml/sec, mL/sec) or 2) Milliliters per minute (ml/min, mL/min); then systems wi" Recalling firm is another firm in the same product code.
first seen 2026-07-08 · recall res_event_number:95254
Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device). Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.