TEX20/TEX20 Pro/TEX20S/TEX20T/TEX20 Exp/TEX20 Elite Diagnostic Ultrasound System, TEX10/ TEX10 Pro/TEX10S/TEX10T/TEX10 Exp/TEX10 Elite/TE X/TE X Lite Diagnostic Ultrasound System

K212265

Shenzhen Mindray Bio-Medical Electronics Co.,LTD · cleared 2021-11-16 · product code IYN · Radiology

Premarket evidence — what FDA accepted

Device typehardware with ml
source quote (p.5)
The TEX20/TEX20 Pro/TEX20S/TEX20T/TEX20 Exp/TEX20 Elite/TEX10/ TEX10 Pro/TEX10S/TEX10T/TEX10 Exp/TEX10 Elite/TE X/TE X Lite Diagnostic Ultrasound System is a general purpose, mobile, software controlled, ultrasonic diagnostic system. Its function is to acquire and display ultrasound images in Modes of operation include: B, M, PWD, CWD, Color Doppler, Amplitude Doppler, Combined mode(B+M, PW+B, Color+B, Power+B, PW+Color+B, Power+PW+B), Tissue Harmonic Imaging, Smart3D, iScape View, TDI, Color M, Strain Elastography, Contrast imaging (Contrast agent for LVO), and Contrast imaging (Contrast agent for Liver). It is an AI (artificial intelligence) based feature. It is used to recognize the interscalene and supraclavicular brachial plexus. It is intended for visualization purpose only and not intended for live guidance during interventional procedure.
AlgorithmAI (artificial intelligence) based feature used to recognize the interscalene and supraclavicular brachial plexus for visualization purpose only.
source quote (p.8)
It is an AI (artificial intelligence) based feature. It is used to recognize the interscalene and supraclavicular brachial plexus. It is intended for visualization purpose only and not intended for live guidance during interventional procedure.
Adaptive (vs locked)FDA source did not state this
PCCPFDA source did not state this
Cybersecurity addressedYes
source quote (p.8)
ClamAV It is an antivirus software

Validation studies (1)

Bench

sample size not stated

endpoints: acoustic output; biocompatibility; cleaning and disinfection effectiveness; thermal safety; electrical safety; mechanical safety

standards: NEMA UD 2-2004 (R2009), AAMI / ANSI ES60601-1:2005/(R)2012 and A1:2012 c1:2009/(r)2012 and a2:2010/(r)2012 (consolidated text), IEC 60601-1-2 Edition 4.0 2014-02, IEC 60601-1-6 Edition 3.1 2013-10, IEC 60601-2-37 Edition 2.1 2015, ISO 14971 Second edition 2007-03-01, ISO 10993-1 Fifth edition 2018-08, IEC 62304 Edition 1.1 2015-06, IEC 62366-1 Edition 1.0 2015-02

Reported performance (0 observations)

FDA source did not state a quantitative performance metric — non-reporting is itself the signal.

Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.

Predicate network

Postmarket — what happened after clearance

85
recalls in product code, 24mo
554
MAUDE reports in code, 12mo
+85%
vs code's own 3-yr baseline
4
drift signals on this device
  • recall_reason_pattern

    Software/algorithm-related recall in product code IYN (Philips Ultrasound, LLC, initiated 2025-10-31): "Ultrasound system compatibility issues with Apple devices running iOS 18 may cause a failure to perform live imagining." Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:97843

  • recall_reason_pattern

    Software/algorithm-related recall in product code IYN (GE Medical Systems, LLC, initiated 2025-09-18): "The Ultrasound-Guided Attenuation Parameter (UGAP) measurement data may display inaccurate values representing liver steatosis. This could potentially lead to inappropriate clinica" Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:97726

  • recall_reason_pattern

    Software/algorithm-related recall in product code IYN (GE Medical Systems China Co., Ltd. Dev. Zone National Hi-Tech; No., initiated 2025-05-16): "GE HealthCare has become aware that the Estimated Fetal Weight (EFW) measurement data feature on the Versana Premier R3 and LOGIQ F R3 series ultrasound systems can display previou" Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:96992

  • recall_reason_pattern

    Software/algorithm-related recall in product code IYN (Siemens Medical Solutions USA, Inc., initiated 2024-08-15): "If ultrasound systems with software, are changed from factory default to : 1) Milliliters per second (ml/sec, mL/sec) or 2) Milliliters per minute (ml/min, mL/min); then systems wi" Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:95254

Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device). Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.

Reimbursement — how devices like this got paid

Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).

RIGOR™ Precedent · public FDA/CMS data · descriptive decision-support, not regulatory or reimbursement advice. Share this page: radar.healthai.com/precedent/device/K212265