Kosmos

K212100

EchoNous, Inc. · cleared 2021-08-06 · product code IYN · Radiology

Premarket evidence — what FDA accepted

Device typesimd
source quote (p.4)
Diagnostic ultrasound system with integrated electronic stethoscope and electrocardiograph
AlgorithmFDA source did not state this
Adaptive (vs locked)FDA source did not state this
PCCPFDA source did not state this
Cybersecurity addressedYes
source quote (p.14)
TIR57:2016 Principles for medical device security - Risk management.

Validation studies (1)

Bench

sample size not stated

standards: CISPR 11:2015+AMD1:2016+AMD2:2019 CSV Consolidated version, ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012, 60601-1-2:2014, 60601-1-6 Edition 3.1 2013-10, 60601-2-37 Edition 2.1 2015, 10993-1:2018, 14971:2019, 62304 Edition 1.1 2015-06 CONSOLIDATED VERSION, 62366-1 Edition 1.0 2015-02, 15223-1 Third Edition 2016-11-01, 62359 Edition 2.1 2017-09 CONSOLIDATED VERSION, UD 2-2004 (R2009), Standard 7351731 Rev. 2.00 2017-02-23, EC53:2013, TIR57:2016, TIR 30:2011

Reported performance (0 observations)

FDA source did not state a quantitative performance metric — non-reporting is itself the signal.

Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.

Predicate network

Postmarket — what happened after clearance

85
recalls in product code, 24mo
554
MAUDE reports in code, 12mo
+85%
vs code's own 3-yr baseline
5
drift signals on this device
  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K233826 (decision 2024-08-29) from EchoNous, Inc. for a matching device line ("Kosmos") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K233826

  • recall_reason_pattern

    Software/algorithm-related recall in product code IYN (Philips Ultrasound, LLC, initiated 2025-10-31): "Ultrasound system compatibility issues with Apple devices running iOS 18 may cause a failure to perform live imagining." Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:97843

  • recall_reason_pattern

    Software/algorithm-related recall in product code IYN (GE Medical Systems, LLC, initiated 2025-09-18): "The Ultrasound-Guided Attenuation Parameter (UGAP) measurement data may display inaccurate values representing liver steatosis. This could potentially lead to inappropriate clinica" Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:97726

  • recall_reason_pattern

    Software/algorithm-related recall in product code IYN (GE Medical Systems China Co., Ltd. Dev. Zone National Hi-Tech; No., initiated 2025-05-16): "GE HealthCare has become aware that the Estimated Fetal Weight (EFW) measurement data feature on the Versana Premier R3 and LOGIQ F R3 series ultrasound systems can display previou" Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:96992

  • recall_reason_pattern

    Software/algorithm-related recall in product code IYN (Siemens Medical Solutions USA, Inc., initiated 2024-08-15): "If ultrasound systems with software, are changed from factory default to : 1) Milliliters per second (ml/sec, mL/sec) or 2) Milliliters per minute (ml/min, mL/min); then systems wi" Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:95254

Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device). Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.

Reimbursement — how devices like this got paid

Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).

RIGOR™ Precedent · public FDA/CMS data · descriptive decision-support, not regulatory or reimbursement advice. Share this page: radar.healthai.com/precedent/device/K212100