AI Segmentation
K203469Varian Medical Systems, Inc. · cleared 2021-04-19 · product code MUJ · Radiology
Premarket evidence — what FDA accepted
source quote (p.4)
“AI segmentation is a web-based application, running in the cloud, that provides a combined deep learning and classical-based approach for automated segmentation of organs at risk, along with tools for structure visualization. This software medical device product is used by trained medical professionals and consists of a web application user interface where the results from the automated segmentation can be reviewed and selected for export into the compatible treatment planning system. Al Segmentation is not intended to provide clinical decisions, medical advice, or evaluations of radiation plans or treatment procedures. Both devices are software-only medical devices.”
source quote (p.4)
“AI segmentation is a web-based application, running in the cloud, that provides a combined deep learning and classical-based approach for automated segmentation of organs at risk, along with tools for structure visualization.”
source quote (p.5)
“Note: These algorithms are static and non-adaptive; they do not alter their behavior over time based on user input.”
Validation studies (1)
Bench
sample size not stated
standards: IEC 62304 Edition 1.1 2015-06 Medical device software - Software life cycle processes, IEC 62366-1 Edition 1.0 2015-02 Application of usability engineering to medical devices, IEC 62083 Edition 2.0 2009-09 Requirements for the safety of radiotherapy treatment planning systems, IEC 82304-1 Edition 1.0 2016-10 Health software - Part 1: General requirements for product safety
Reported performance (0 observations)
FDA source did not state a quantitative performance metric — non-reporting is itself the signal.
Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.
Predicate network
Postmarket — what happened after clearance
- re_clearance
The FDA AI/ML device list shows a newer 510(k) K230082 (decision 2023-05-04) from GE Medical Systems, LLC for a matching device line ("Auto Segmentation") — a new clearance for the same line is a change event.
first seen 2026-07-08 · k_number:K230082
- re_clearance
The FDA AI/ML device list shows a newer 510(k) K211881 (decision 2021-09-02) from Varian Medical Systems, Inc. for a matching device line ("AI Segmentation") — a new clearance for the same line is a change event.
first seen 2026-07-08 · k_number:K211881
- recall_reason_pattern
Software/algorithm-related recall in product code MUJ (Philips Medical Systems (Cleveland) Inc, initiated 2025-08-05): "Due to a software issue, there is a potential image error of the Region of Interest for expansion/contraction for HFP (Head First Prone), FFS (Feet First Supine) and FFP (Feet Firs" Recalling firm is another firm in the same product code.
first seen 2026-07-08 · recall res_event_number:97049
- recall_reason_pattern
Software/algorithm-related recall in product code MUJ (Philips Medical Systems (Cleveland) Inc, initiated 2025-07-17): "Due to software issue, Radiation Therapy Planning system may provide incorrect dataset calculations when performing the "Stopping Power Ratio" (SPR) ," Recalling firm is another firm in the same product code.
first seen 2026-07-08 · recall res_event_number:97309
Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device). Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.
Reimbursement — how devices like this got paid
Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).