Deep Recon
K193073Shanghai United Imaging Healthcare Co., Ltd. · cleared 2020-07-06 · product code JAK · Radiology
Premarket evidence — what FDA accepted
source quote (p.5)
“The Deep Recon is a data driven image reconstruction method based on deep learning technology. ... The function is integrated on the mentioned CT systems as a part of reconstruction chain.”
source quote (p.5)
“The Deep Recon is a data driven image reconstruction method based on deep learning technology. Dedicated deep neural networks are designed and trained for different body parts.”
source quote (p.6)
“UNITED IMAGING HEALTHCARE conforms to the Cybersecurity requirements by implementing a process of preventing unauthorized access, modification, misuse or denial of use, or unauthorized use of information that is stored, accessed, or transferred from a medical device to an external recipient. Cybersecurity information in accordance with guidance document “Content of Premarket Submissions for Management of Cybersecurity in Medical Devices” is included in this submission.”
Validation studies (3)
Bench
sample size not stated
endpoints: Low contrast detectability (LCD); Image noise; Mean CT number and uniformity; Spatial resolution; Reconstructed section thickness
standards: NEMA PS 3.1-3.20(2011): Digital Imaging and Communications in Medicine (DICOM), IEC 62304: Medical Device Software - software life cycle process, ISO 14971: Medical Devices – Application of risk management to medical devices, Code of Federal Regulations, Title 21, Part 820 - Quality System Regulation, Code of Federal Regulations, Title 21, Subchapter J - Radiological Health
Retrospective clinical
n=80 cases
endpoints: image noise; structure fidelity; diagnostic quality
Retrospective clinical
n=40 cases
endpoints: image quality; clinical features; diagnostic quality
Reported performance (0 observations)
FDA source did not state a quantitative performance metric — non-reporting is itself the signal.
Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.
Predicate network
Postmarket — what happened after clearance
- recall_reason_pattern
Software/algorithm-related recall in product code JAK (GE Medical Systems, LLC, initiated 2026-03-26): "GE HealthCare has become aware of a potential security vulnerability impacting AW Server deployed via Edison Health Link (EHL) based CT Smart Subscription used in conjunction with " Recalling firm is another firm in the same product code.
first seen 2026-07-08 · recall res_event_number:98738
- recall_reason_pattern
Software/algorithm-related recall in product code JAK (PHILIPS MEDICAL SYSTEMS, initiated 2026-03-07): "Philips has identified three software issues: 1. During a continuous CT (CCT) scan, there is the potential that the Gantry could remain at the current scan position after pressing" Recalling firm is another firm in the same product code.
first seen 2026-07-08 · recall res_event_number:98588
- recall_reason_pattern
Software/algorithm-related recall in product code JAK (Siemens Medical Solutions USA, Inc, initiated 2025-12-19): "To remove the software applications from certain CT systems as the applications have not received FDA 510(k) clearance." Recalling firm is another firm in the same product code.
first seen 2026-07-08 · recall res_event_number:98206
- recall_reason_pattern
Software/algorithm-related recall in product code JAK (PHILIPS MEDICAL SYSTEMS, initiated 2025-09-25): "Issue 1: The potential for unintentional continued gantry/couch movement when a specific button series is used requiring use of manual stop. Issue 2. When performing a helical/Axia" Recalling firm is another firm in the same product code.
first seen 2026-07-08 · recall res_event_number:97699
- recall_reason_pattern
Software/algorithm-related recall in product code JAK (Philips North America Llc, initiated 2025-05-29): "Devices with affected software may experience two unintended motion issues that may lead to contact between the Gantry or table with the operator or patient, along with additional " Recalling firm is another firm in the same product code.
first seen 2026-07-08 · recall res_event_number:97010
- recall_reason_pattern
Software/algorithm-related recall in product code JAK (, initiated 2025-05-23): "Due to a software bug the robotic arm may incorrectly guide the instrument compared to the planning defined by user." Recalling firm is another firm in the same product code.
first seen 2026-07-08 · recall res_event_number:96967
- …and 3 more.
Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device). Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.