BriefCase
K180647Aidoc Medical, Ltd. · cleared 2018-08-01 · product code QAS · Radiology
Premarket evidence — what FDA accepted
source quote (p.4)
“BriefCase is a radiological computer-assisted triage and notification software device. The software system is based on an algorithm programmed component and is comprised of a standard off-the-shelf operating system, the Microsoft Windows server 2012 64bit, and additional applications, which include PostgreSQL, DICOM module and the BriefCase Image Processing Application.”
source quote (p.3)
“BriefCase uses an artificial intelligence algorithm to analyze images and highlight cases with detected ICH on a standalone desktop application in parallel to the ongoing standard of care image interpretation. Specifically, the subject and predicate software utilize a deep learning algorithm trained on medical images.”
Validation studies (1)
Retrospective clinical
n=198 cases · 3 site(s)
endpoints: software's performance in identifying non-contrast CT head images containing intracranial hemorrhage (ICH) findings; Briefcase's potential clinical benefit of worklist prioritization for true positive ICH cases
Reported performance (4 observations)
source quote (p.7)
“Specifically, sensitivity was observed to be 93.6% (95% CI: 86.6%-97.6%)”
source quote (p.7)
“and specificity was observed to be 92.3% (95% CI: 85.4%-96.6%).”
source quote (p.8)
“Time-to-open-exam in the standard of care 72.58 45.02 100.14 41.00”
source quote (p.8)
“Time-to-notification of BriefCase 4.46 4.10 4.83 3.95”
Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.
Predicate network
Postmarket — what happened after clearance
- re_clearance
The FDA AI/ML device list shows a newer 510(k) K253578 (decision 2026-02-26) from Aidoc Medical , Ltd. for a matching device line ("BriefCase-Triage: CARE Multi-Triage CT for Pneumothorax; Pericardial effusion; Large aortic aneurysm; Shoulder fracture or dislocation device") — a new clearance for the same line is a change event.
first seen 2026-07-08 · k_number:K253578
- re_clearance
The FDA AI/ML device list shows a newer 510(k) K251195 (decision 2026-01-27) from Aidoc Medical , Ltd. for a matching device line ("BriefCase-Triage") — a new clearance for the same line is a change event.
first seen 2026-07-08 · k_number:K251195
- re_clearance
The FDA AI/ML device list shows a newer 510(k) K252970 (decision 2026-01-07) from Aidoc Medical , Ltd. for a matching device line ("BriefCase-Triage: CARE Multi-triage CT Body") — a new clearance for the same line is a change event.
first seen 2026-07-08 · k_number:K252970
- re_clearance
The FDA AI/ML device list shows a newer 510(k) K253265 (decision 2025-11-06) from Aidoc Medical , Ltd. for a matching device line ("BriefCase-Triage") — a new clearance for the same line is a change event.
first seen 2026-07-08 · k_number:K253265
- re_clearance
The FDA AI/ML device list shows a newer 510(k) K251406 (decision 2025-05-30) from Aidoc Medical, Ltd. for a matching device line ("BriefCase-Triage") — a new clearance for the same line is a change event.
first seen 2026-07-08 · k_number:K251406
- re_clearance
The FDA AI/ML device list shows a newer 510(k) K250248 (decision 2025-02-14) from Aidoc Medical, Ltd. for a matching device line ("BriefCase-Triage") — a new clearance for the same line is a change event.
first seen 2026-07-08 · k_number:K250248
- …and 27 more.
Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device). Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.
Reimbursement — how devices like this got paid
Not yet tracked — no payment pathway indexed for this clearance (the reimbursement corpus is a growing seed set).