Lung Vision

K163622

BODY VISION MEDICAL LTD. · cleared 2017-05-11 · product code LLZ · Radiology

Premarket evidence — what FDA accepted

Device typesimd
source quote (p.4)
LungVision image processing algorithms are executed on a PC based hardware platform, which can perform the following functions: segment previously acquired DICOM 3D CT image data, register DICOM 3D CT image data with live fluoroscopic X-ray image overlay the segmented 3D CT dataset over a live fluoroscopic X-ray image of the same anatomy, obtained on a Fluoroscopic system.
AlgorithmImage processing algorithms for segmenting 3D CT data, registering 3D CT data with live fluoroscopic X-ray images, and overlaying segmented 3D CT datasets over live fluoroscopic X-ray images.
source quote (p.4)
LungVision image processing algorithms are executed on a PC based hardware platform, which can perform the following functions: segment previously acquired DICOM 3D CT image data, register DICOM 3D CT image data with live fluoroscopic X-ray image overlay the segmented 3D CT dataset over a live fluoroscopic X-ray image of the same anatomy, obtained on a Fluoroscopic system.
Adaptive (vs locked)FDA source did not state this
PCCPFDA source did not state this
Cybersecurity addressedNo

Validation studies (1)

Bench

n=1 other

endpoints: met all requirements specifications; verification testing of the requirements; testing of hazards mitigations; performance testing of the system; accuracy in deformable tissue

standards: IEC 60601-1:2005, IEC 60601-1-2:2014, ANSI AAMI ES 60601-1, IEC 60601-1-2

Reported performance (0 observations)

FDA source did not state a quantitative performance metric — non-reporting is itself the signal.

Each value carries its own analysis unit and task — never compare or pool across devices. Source: 510(k) summary PDF.

Predicate network

Postmarket — what happened after clearance

48
recalls in product code, 24mo
295
MAUDE reports in code, 12mo
+683%
vs code's own 3-yr baseline
27
drift signals on this device
  • re_clearance

    The FDA AI/ML device list shows a newer 510(k) K183593 (decision 2019-04-18) from Body Vision Medical Ltd. for a matching device line ("Lung Vision System") — a new clearance for the same line is a change event.

    first seen 2026-07-08 · k_number:K183593

  • adverse_event_inflection

    MAUDE adverse-event reports for product code LLZ: 295 in the 12 months ending 2026-06, vs a 37.7/12mo average over the prior 3 windows (+683%). Code-level count — reports are not attributed to this specific device.

    first seen 2026-07-08 · openFDA /device/event.json count=date_received product_code=LLZ

  • recall_reason_pattern

    Software/algorithm-related recall in product code LLZ (GE Medical Systems SCS, initiated 2026-05-08): "GE HealthCare has become aware of a context synchronization issue in AW Server 3.2 ext. 6.5. When a user selects a patient or exam in the AW Server Web Client worklist and launches" Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:99042

  • recall_reason_pattern

    Software/algorithm-related recall in product code LLZ (PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. Veenpluis 4-6 Best Netherlands, initiated 2026-03-05): "Potential that mis-ordered frames in Vue Motion during dynamic cine runs may cause images frames to display out of sequence." Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:98520

  • recall_reason_pattern

    Software/algorithm-related recall in product code LLZ (GE Medical Systems, LLC, initiated 2026-01-30): "There is a potential cybersecurity vulnerability affecting certain versions of Centricity Universal Viewer. User login credentials may be exposed on the local client workstation, w" Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:98428

  • recall_reason_pattern

    Software/algorithm-related recall in product code LLZ (Focalyx Technologies, LLC., initiated 2025-12-23): "Software device that is an accessory for image-guided interventional and diagnostic procedures involving prostate has accessories that may not function properly with Windows 10, wh" Recalling firm is another firm in the same product code.

    first seen 2026-07-08 · recall res_event_number:98137

  • …and 21 more.

Recall and MAUDE counts are product-code-level (reports aren't reliably attributable to one device). Signals are descriptive observables with sources — never a judgment that the device is unsafe or drifting. Snapshot 2026-07-08.